5-3-5. Numeric Value Registering Calibration
The numeric value registering calibration calibrates data by changing the initial value.
Execute the numeric value registering calibration after the actual load calibration and sensitivity registering calibration.
The numeric value registering calibration is used for pressure transducers for absolute pressure measurement or when
already load is applied and unable to obtain the actual load state. By registering the initial value of the transducer
written in the Test Data Sheet or registering an already measured initial value, a reference point equivalent to unload
state can be registered.
<Setting Range>
-3.2000 to +3.2000 mV/V
<How to operate>
1) Press the
Next
2) Press the
Numeric Regist Cal.
3) The initial value, registered during the no-load zero, appears.
In the case of this example, the "0.1000 mV/V" appears.
4) Enter the new initial value with the numeric keyboard.
In the case of this example, as you enter "-0.5432", the "-
0.5432mV/V" appears.
5) Press the
Register
The "Registering ini Val." window blinks on the top left for
registering the new initial value.
key on the "Calibration (1/2)" window.
key on the "Calibration (2/2)" window.
key.
Calibration
Calibration
Numeric Regist Cal.
Numeric Reg.Cal.
ini.Val.
New
+/-
Input the desired value with
keyboard.Push 'Register'key.
Numeric Reg.Cal.
ini.Val.
New
+/-
Input the desired value with
keyboard.Push 'Register'key.
Registering
ini. Val.
ini.Val.
Blinks
New
+/-
Input the desired value with
keyboard.Push 'Register'key.
47
Return
'Bridge Voltage','No-Load Zero
Manual Input Cal.
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Calibration is automatically
TEDS Auto Cal.
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
TEDS Part Cal.
the only selected item from
TEDS built-in sensor.
'TEDS Read Ope.'and 'Zero dur-
TEDS Ope.Setting
ing TEDS'are set.TEDS informa-
tions are displayed.
Next
Return
Change initial value. Run
after 'Actual Load Cal.' or
'Sens.Regist.Cal.'.
Previous
Return
Now
0.1000 mV/V
mV/V
7
8
9
C
.
4
5
6
0
1
2
3
Return
Now
0.1000 mV/V
-0.5432 mV/V
7
8
9
C
.
4
5
6
0
1
2
3
Registering
Return
ini. val.
Now
0.1000 mV/V
-0.5432 mV/V
7
8
9
C
.
4
5
6
0
1
2
3
Measure
1 / 2
Measure
2 / 2
Register
Register
Register