Download Print this page

KYOWA WGA-910A Series Instruction Manual page 61

Hide thumbs Also See for WGA-910A Series:

Advertisement

6-2-2. Zero Compensation
Automatically corrects subtle zero point drifts.
It effectively reduces measurement error inducible to zero drifts on a sensor or on the WGA-910A as well as zero
drifts caused by deposits on a measuring object.
(1) ZERO every Comp. Time (Zero Tracking)
Conducts digital zero and automatically sets the display value to "0" when the measured value is within the
compensation range (tracking band) at every determination time (tracking delay) passes.
The additional value is displayed when it is set.
<Setting Range>
J.Time (Judging Time):
C.Range (Compensation Range): 0 to 99999 (count)
<How to operate>
1) Press the
2) Press the
3) Press the
4) Press the
window.
5) After pressing the
numeric keyboard and press the
value.
6) After pressing the
numeric keyboard and press the
value.
MEMO
The compensation range is + and - sides of 0.
The "ZERO every Comp. Time" is not conducted when the "J.
Time" is set to 0.00 sec.
0.00 to 9.99 (sec)
key on the "Measuring" window.
key on the "Setting" window.
key on the "Input Setting" window.
key on the "Zero Compensation"
key, input a setting value with the
key to register the setting
key, input a setting value with the
key to register the setting
57
Input Setting
Return
A function that stables the
Smoothing
indicated value by reducing
unnecessary noise.
A function that automatically
Zero Compensation
compensates the fine zero-
shift.
The measured value is display-
Additional Value
ed with adding the 'Additional
Value'.
The original value of sensor
Original Value
is monitored.
Measure

Advertisement

loading

This manual is also suitable for:

Wga-910a-0Wga-910a-1Wga-910a-2Wga-910a-3Wga-910a-4Wga-910a-12 ... Show all