5-3-3. TEDS Auto Cal. (TEDS-based Automatic Calibration)
The calibration is automatically conducted by reading the required information for the calibration (applied voltage,
rated output, and unit) from the TEDS built-in sensor and the calibration data that is saved inside is overwritten.
However, the "TEDS-based Automatic Calibration" is valid when "Allow" is selected in the "Key Ops." of the
"TEDS Reading Operation" window.
The
In addition, the TEDS information is read after conducting the no-load zero and calibration is conducted when
selecting "Execute" in the "Zero during TEDS" window.
<How to operate>
1) Press the
2) After pressing the
Cal." window appears.
3) Press the
information is read and the calibration is started.
Press the
previous window.
4) The calibration starts and the "TEDS Calibration" window on the
upper left blinks.
5) The calibration is finished after a short time and a new calibration
value is displayed on the "New" column.
Press the
key is invalid and displayed in grey when "Prohibit" is selected.
key on the "Calibration" window.
key, the "Confirmation of TEDS Auto
key. After conducting the no-load zero, the TEDS
key to cancel the calibration and gets back to the
key at last to move to the "Measuring" window.
Blink
43
Calibration
Return
'Bridge Voltage','No-Load Zero
Manual Input Cal.
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Calibration is automatically
TEDS Auto Cal.
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
TEDS Part Cal.
the only selected item from
TEDS built-in sensor.
'TEDS Read Ope.'and 'Zero dur-
TEDS Ope.Setting
ing TEDS'are set.TEDS informa-
tions are displayed.
Next
Measure
1 / 2