3-3. WIRING
3-3-1. Connecting transducer
・Input connector is NDIS one-touch connector.
・When using a KYOWA's transducer with the NDIS standard connector at the tip of the cable, the both TEDS-
compatible and non-TEDS-compatible sensors (excluding six-wire system remote sensing) can be directly
connected to an input connector
・When using the KYOWA's sensor without a connector at the tip of the cable or when using other manufactures
made sensor, connect the sensor to the NDIS standard connector (option) as follows.
Applicable connector
Pin No.
A
B
C
D
E
F
G
MEMO
Cable colors described above are
KYOWA's standard color codes.
NOTE 1
・For transducer wiring, use shortest possible 4-conductor or 6-conductotor shielded cable.
In addition, locate it apart from the power cable or other wirings that may generate noise.
・Bridge excitation voltage is set to 2 V prior to shipment. Since it can be selected to 10 V, set the
voltage appropriate for sensor specifications (within the recommended excitation voltage).
For details, see "5-3-2. Manual Input Calibration"
NOTE 2
・Never connect a 6-wire system remote sensing sensor.
Or, the WGA-910A may cause trouble or accident.
・When connecting a non-TEDS-compatible sensor, don't connect the F and G terminals of the NDIS
standard connector.
Model:
Manufacture:
Signal Name
+BV (Red)
-SIG (White)
-BV (Black)
+SIG (Green)
Shield
TEDS(+) (Yellow)
TEDS(-) (Blue)
PRC03-12A10-7M10
TAJIMI ELECTRONICS CO., LTD.
21