6. OPERATION AND FUNCTION
6-1. CALIBRATION
For the "Manual Input Calibration," "TEDS-based Automatic Calibration," and "Partial Calibration Based on TEDS
Calibration Data," refer to "5. CALIBRATION."
This chapter describes how to operate the digital zero function on the "Measuring" window and how to set for the
TEDS operation.
6-1-1. Digital Zero Function
Sets the current displaying value to "0 (additional value when setting the value to other than 0)."
This is pretty much same as resetting a pointer to '0' before you puts an object on a scale.
The digital zero function has 2 ways, by key operation and by control input operation from the rear panel.
(1) By key operation
1) Press
2) The "Confirmation of Digital Zero" window appears.
Press the
becomes zero.
An additional value is displayed when the value is set to other than
zero.
Press the
value gets back to the measured value that is based on the no-load
zero.
Press the
the measured value.
NOTE
The zero point value that is obtained by the key operation is
maintained even if the WGA-910A is turned OFF.
3) After conducting the digital zero and gets back to the "Measuring"
window, the
MEMO
Yellow key: Conducting digital zero.
Gray key:
key on the "Measuring" window.
key to conduct digital zero and the display value
key to release the digital zero and the display
key to cancel the digital zero and gets back to
key changes to yellow.
Digital zero is released.