(4) Supplemental explanations of the TEDS function
(1) Executing the TEDS calibration at the initial value.
Set the sensor in no-load state and conduct the calibration with the TEDS. The no-load zero is automatically
conducted and the display value is set to "0 (or additional value)" when the Execute is selected in the "ZERO
during TEDS" window.
Then automatically conducts the sensitivity registering calibration, updates calibration values of the excitation
voltage, rated output, rated display (including a decimal digit position), and unit.
Capable of executing the TEDS calibration by the key operation on the window, control input, when turning ON
the power and by communication command.
However, executing no-load zero to the initial value when turning ON the power is prohibited.
MEMO
For replacing the sensor to the same model and same capacity one because of the sensor's faulty, set the Rd.
Out. of the TEDS Cal. Item in the "TEDS Part Cal." window to "Execute" and Ex. Volt, Rd. Cap. and
Unit to "Prohibit" for the easy calibration.
(2) Applied voltage
The applied voltage should be equal or less to the recommended applied voltage and higher one should be
selected.
[Example]
(3) Calculation of the calibration coefficient
Register the calibration coefficient that is obtained from the following expression.
Rated capacity of the sensor (y) ÷ Rated output (x)
(4) No-load zero
No-load zero of the TEDS calibration is same with that of the manual input calibration.
(5) Display value and unit
Decimal point of the rated capacity (display) and unit after conducting the TEDS calibration is automatically set
corresponding to the total number of display digits of 5.
[Example]
(6) Multiple connection of the TEDS-compatible sensor
Combined rated capacity and rated output are calculated as follows when connecting the multiple TEDS-
compatible sensors (2 to 4) in parallel.
Rated Capacity (combined): Total value of the rated capacities of the sensors.
Rated Output (combined): Average value of the rated outputs of the sensors (Value obtained by dividing the
total value of sensors with the number of units.).
(7) Model, serial No., and calibrated date of the TEDS
For reading the KYOWA made sensor only.
The information may not be read when connecting the other made TEDS-compatible sensor.
2 V is set for the recommended applied voltage of 3 V.
5 N = 5.0000 N, 500 N = 500.00 N, 50 N = 50.000 N, 5 kN = 5.0000 kN
52