Download Print this page

Model - KYOWA WGA-910A Series Instruction Manual

Hide thumbs Also See for WGA-910A Series:

Advertisement

3-3-2. Connecting Control Input
・Applicable connector of the cable side is as follows.
Connector

model:

Case model:
・Pin numbers and signal names of the control input are as follows.
Pin No.
19
ZERO command
20
HOLD command
21
RESET command
22
TEDS command
23
Waveform command
24
Measuring condition select 0
25
Measuring condition select 1
26
Measuring condition select 2
27
Measuring condition select 3
・When using external switch, open collector, etc. to control inputs, connect as shown in the following figure.
NOTE
・Never connect the signals to reserve pin.
・The external element of control input requires 10 mA or more current.
・Leakage current on the external element of control input must be below 50 uA.
・A voltage on the external element of control input must be below 2 V when it is put [ON] (short-
circuited)connector.
NOTE for wiring
・For connecting more than 1-m long cable, use a twisted cable.
・For connecting more than 3-m long cable, use a shielded cable and always ground the shielded cable
to the ground terminal.
・In addition, for connecting more than 30-m long cable or using outdoors, take special countermeasures
against noise.
PCR-S36FS+
PCR-LS36LA
Signal Name
Manufacture: HONDA TSUSHIN KOGYO CO., LTD.
Manufacture: HONDA TSUSHIN KOGYO CO., LTD.
Pin No.
28
Reserve
29
Reserve
30
Reserve
31
Reserve
32
Reserve
33
Reserve
34
Reserve
35
Input COM
36
Input COM
22
Signal Name