(7)[GST]
Content:
Format:
Parameter
Return value format:
Description and others: Return value abcdefghi denotes the status with '1' or '0.'
(8)[GOD]
Content:
Format:
Parameter:
Return value format:
(9) [PRR]
Content:
Format:
Parameter
Return value format:
Read the status.
GST<CR>
None
Sabcdefghi<CR><LF>
Numeric '1' is output in the following state.
a: HH comparator is ON.
c: OK comparator is ON.
e: LL comparator is ON.
g: Detecting.
i: Error
Display the original value (sensor output value).
GOD<CR>
None
DO±x.xxxx<CR><LF>
Read the setting value.
PRRppp<CR>
ppp
010: Sensitivity registering value
011: Numeric Value Registering Calibration value
012: Excitation voltage
030: Additional value
051: Compared value of the HH comparator
052: Compared value of the HI comparator
053: Compared value of the LO comparator
054: Compared value of the LL comparator
058: Measuring condition No.
090: ZERO compensation function
115: Keylock
131: Waveform display setting 2
150: Measuring condition selection signal
151: Backlight lightning time
152: Language
For details, see each command. (Excluding the options.)
000<CR><LF>
Option: None (WGA-910A-0)
b: HI comparator is ON.
d: LO comparator is ON.
f: During measurement.
h: During holding.
Sensor output value: -3.2000 to 3.2000 (mV/V)
020: Smoothing function
041: Comparison setting
070: Setting measuring mode
100: Option
130: Waveform display setting 1
153: Clock
119