(3-b) Actual Load Cal. (Actual Load Calibration)
Actual load is applied to the sensor for the calibration. Displays the output when an actual load is applied as the
rated output. Suppose that you are using a transducer with the "Rated Capacity: 200 kN" and "Rated Output:
2.0040 mV/V." In order to read "200.00" against a load of 200 kN that you will give to the transducer, do as
follows.
<Setting range>
Rated Output: (-3.2000 to +3.2000 mV/V; display only, excluding 0)
Rated Display: -99999 to 99999 (count; excluding 0)
<How to operate>
1) When a load is applied to the sensor, the output is displayed to the
"New" column of the rated output.
Check that the value is stabled.
[2.0040 mV/V] is displayed in the right figure.
2) Input a setting value of the "Rated Output" with the numeric
keyboard.
Suppose "200.00" is input, "200.00 kN" is displayed in the "New"
column.
3) Press the
upper left blinks and the "Rated Output" and "Rated Display" are
registered.
4) After the registration, press the
window.
MEMO
The "Sens. Registering" window blinks when conducting the actual load calibration with communication
during the measurement or during the setting.
key. The "Sens. Registering" window on the
key to move to the "Unit"
Sens.Regis-
Sensitivity reg-
istering
tering
Calibration
Input Setting
Meas.Cond.Setting
Wave Display Setting
41
Blink
Measure
Calibration is conducted with
sensor by 'Manual Input','TEDS
Auto'or'TEDS Part'Cal.
'Smoothing','Additional Value'
and 'Zero Compensation'are set
The sensor output is monitored
'Comparison Set'and 'Measure
Mode Set'are set as measure-
ment condition.
The necessary setting to dis-
play the wave such as 'X-axis'
and 'Y-axis'are set.
Next
1 / 2

Advertisement

loading

This manual is also suitable for:

Wga-910a-0Wga-910a-1Wga-910a-2Wga-910a-3Wga-910a-4Wga-910a-12 ... Show all