5-3-4. TEDS Part Cal. (Partial Calibration Based on TEDS Calibration Data)
Only the selected information from the TEDS built-in sensor (select one or multiples from the applied voltage, rated
display, rated output, unit) is updated the calibration data that was saved internally and calibrated automatically.
However, the "Partial Calibration Based on TEDS Calibration Data" is valid when selecting the "Key Ope." in
the "TEDS Reading Operation" to "Allow."
The
In addition, the TEDS information is read and calibration is conducted after conducting the no-load zero when
setting the "Zero during TEDS" to "Execute."
Select the target item for updating in the "TEDS Cal. Item" window.
<Setting Range>
(Common to the Ex. Volt., Rd. Out., Rd. Cap., and Unit)
<How to operate>
1) Press the
2) Press the
3) Every time the right key of the calibration item is pressed,
or
changes.
After the setting, press the
After the registration, press the
4) Press the
Cal." window.
MEMO
Calibration item with the "Prohibit" selected in the "TEDS Cal.
Item" window, "New" column is not displayed in the "TEDS Part
Cal." Window.
5) Press the
information is read and the calibration starts.
Press the
previous window without conducting the calibration.
key is invalid and displayed in gray when setting "Key Ope." to "Prohibit."
key on the "Calibration" window.
key on the "TEDS Part Cal." window.
key.
key.
key to display the "Confirmation of TEDS Part
key. After conducting the no-load zero, the TEDS
key to cancel the calibration and gets back to the
Execute or Prohibit
45
Calibration
Return
'Bridge Voltage','No-Load Zero
Manual Input Cal.
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Calibration is automatically
TEDS Auto Cal.
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
TEDS Part Cal.
the only selected item from
TEDS built-in sensor.
'TEDS Read Ope.'and 'Zero dur-
TEDS Ope.Setting
ing TEDS'are set.TEDS informa-
tions are displayed.
Next
Measure