6-4-3. Start Mode of Wave (Trigger Mode)
Sets a start mode (trigger mode) for displaying the waveform. Operations of modes are described as follows.
<Setting Range>
Auto Mode, Normal Mode, Single Mode
<How to operate>
1) Press the
window.
2) Press the target key and
(1) Auto Mode
Starts displaying the waveform and automatically updates it even if the trigger conditions are not satisfied.
Press the
display.
After displaying the waveform of one window to the end point of the X-axis, the waveform is held in the
"Waveform Holding Time" and then the next waveform is displayed repeatedly.
Press the
the "Waveform Command" for stop
loading.
MEMO
Satisfying trigger condition means starts
displaying the waveform when the
measured value passed the preset level in
the Passed Level and the measured
value passed as presetting in the Level
Passed Way.
(2) Normal Mode
Press the
The WGA-910A becomes trigger condition waiting mode (see the above MEMO) and starts displaying the
waveform only when the measured value satisfies the trigger condition.
After displaying the waveform of
one window until the end point of
the X-axis, the WGA-910A becomes
trigger condition waiting mode and
holds the waveform without updating
it.
key on the "Wave Display Setting (1/2)"
key for registration.
key or turn ON the "Waveform Command" of the control input for starting the waveform
key or turn OFF
key or turn ON the "Waveform Command."
82
Wave Display Setting
Return
The end point of X-axis(time)
X-axis Setting
is set.
10.0sec
The start and end point of Y-
Y-axis Setting
axis are set.
1000.00kN
The mode to start the wave
Start Mode of Wave
display is set.
Auto Mode
The level that measured value
Passed Level
passes for satisfying trigger
condition is set.
0.00kN
Next
Measure
1 / 2