KYOWA WGA-910A Series Instruction Manual page 50

Hide thumbs Also See for WGA-910A Series:

Advertisement

6) The calibration starts and the "TEDS Calibration" window on the
upper left blinks.
7) The calibration finishes after a short time and a new calibration value
is displayed on the "New" column.
Press the
key at last to get back to the "Measuring" window.
NOTE
"TEDS Error" window blinks when the TEDS information cannot read correctly or TEDS information
value is out of specifications.
After stopping the reading of the TEDS information, the "TEDS Error" window blinks approx. 3 seconds.
After that, the current calibration value is displayed.
At this time, the calibration value is not updated.
MEMO
The "TEDS Calibration" window blinks after the following settings or conducting the following operations
during the setting or during the setting.
<Setting>
Set "Cnt. Input", "Power ON", "Command" to "Allow" in the "TEDS Reading Operation" window.
<Operation>
・TEDS command of the control input is turned ON.
・Receives a communication command (EID) that executes the automatic sensitivity-registering calibration
with the TEDS.
・After turning ON the power again.
Blink
46