Calibration of the Touch Screen

46
When E5061B measurement screen appears, you have to
calibrate the touch screen. Follow the procedure described
below to calibrate the touch screen.
Step 1. Press
Step 2.
Press Service Menu.
Step 3.
Press Test Menu.
Step 4.
Press Adjust Touch Screen.
The touch screen calibration screen
Figure 18 Touch Panel Calibration Screen
Step 5.
Touch the symbol on the upper left of the screen
with your finger. The symbols also appears on the upper
right, lower left, and lower right of the screen. Touch all the
symbols in that order with your finger.
Touching all the four locations described above with your
finger automatically concludes the touch screen calibration.
.
(Figure
18) appears.