Removing The Source Enclosure From The Mass Spectrometer - Waters TRIZAIC UPLC System Manual

Table of Contents

Advertisement

To prepare the instrument:
1.
Ensure the mass spectrometer is in Standby mode.
For a Xevo TQ, in the MassLynx Tune window, click Standby
For mass spectrometers other than the Xevo TQ, in the MassLynx
Tune window, click Source Standby
2.
Set the source temperature to 30 °C.
3.
Wait 30 minutes, to allow the source to cool.
The API gas flow stops automatically when you click Source
Note:
Standby.
Exception:
the API gas flow.
The relevant instrument's operator's guide.
See also:

Removing the source enclosure from the mass spectrometer

Requirements:
You must remove the currently fitted source enclosure from the mass
spectrometer before installing the TRIZAIC UPLC source enclosure.
For a SYNAPT mass spectrometer, you must install the source adaptor
housing after removing the SYNAPT source enclosure (see
Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
Source adaptor housing
2-2
Preparing the System
For a Xevo TQ, in the Tune window, click API
.
to stop
page
2-4).
.