Starting The Mass Spectrometer - Waters Xevo TQD Overview And Maintenance Manual

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

2 Preparing for Operation

Starting the mass spectrometer

The Waters Xevo TQD is compatible with the ACQUITY UPLC system; if you
are not using an ACQUITY UPLC system, refer to the documentation relevant
to the system you are using (see "Non-ACQUITY devices for use with Xevo
TQD" on
page
Notice:
compatible solvents only. For more details, refer to the following
sources:
• See
• Appendix C of the ACQUITY UPLC System Operator's Guide for
solvent compatibility with ACQUITY devices.
Starting the mass spectrometer entails powering-on the workstation, logging
in, powering-on the mass spectrometer and all other instruments, and then
starting the MassLynx or UNIFI software.
Requirement:
that it obtains the IP addresses of the system instruments.
See the mass spectrometer's online help for details on MassLynx, IntelliStart,
and UNIFI software.
To start the mass spectrometer:
Warning:
supply pressure fall below 400 kPa (4.0 bar, 58 psi).
1.
On the rear panel, ensure the nitrogen supply is connected to the
instrument's nitrogen inlet connection (see the figure on
Requirement:
least 95% or, for APGC use, at least 99.999%. Regulate the supply at 600
to 690 kPa (6.0 to 6.9 bar, 90 to 100 psi).
For more information on connections, see the figure on
2.
Ensure that the collision gas supply is connected to the instrument's
collison cell gas inlet.
Requirement:
purity (99.997%). Regulate the supply at 50 kPa (0.5 bar, 7 psi).
46
March 14, 2014, 715004387 Rev. B
29).
To avoid causing severe damage to the instrument, use
page 243
for solvent information.
You must power-on and log in to the workstation first to ensure
To avoid igniting flammable solvents, never let the nitrogen
The nitrogen must be dry and oil-free, with a purity of at
The collision gas is argon; it must be dry and of high
page
44).
page
44.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents