Typical Tests To Perform; Frequency Of Tests - GE LOGIQ V2 User Manual

Hide thumbs Also See for LOGIQ V2:
Table of Contents

Advertisement

After the Exam is Over

Typical Tests to Perform

Frequency of tests

3-54
Quality assurance measurements provide results relating to
system performance. Typically these are:
Axial Measurement Accuracy
Lateral Measurement Accuracy
Axial and Lateral Resolution
Penetration
Functional & Contrast Resolution
Gray Scale Photography.
With these tests, a performance baseline can be set at
installation with the phantom in your department. Future test
results can be compared to the baseline in order to maintain a
record of system performance trends.
Quality assurance tests are used to determine whether a
scanner is providing the same level of performance from day to
day.
The frequency of testing varies with the amount of system usage
and modes to be tested. It is recommended that the user
perform quality assurance tests at least every three months or
every 400 patient studies. Tests should also be performed when
a question about system performance exists.
A mobile system may require more frequent tests.
Image quality should also be tested immediately after the
following events:
Service calls
System upgrades/modifications
Dropped probe, power surge, etc.
LOGIQ V2/LOGIQ V1
Direction 5610736-100 Rev. 9
User Guide