3.7 Electrical Connections

For safe connections, no uninsulated wiring should be able to come
in contact with fingers, tools or clothing during normal operation.
CAUTION:
USE SHIELDED CABLES. ALSO, USE PLUGS THAT
PROVIDE EXCELLENT EMI/RFI PROTECTION. THE
PLUG CASE MUST BE CONNECTED TO THE CABLE
SHIELD, AND IT MUST BE TIGHTLY FASTENED TO
THE ANALYZER WITH ITS FASTENING SCREWS.
ULTIMATELY, IT IS THE INSTALLER WHO ENSURES
THAT THE CONNECTIONS PROVIDE ADEQUATE
EMI/RFI SHIELDING.
3.7.1
24VDC I
NPUT
The analyzer is powered using a 24 VDC source. Insert the power
supply connector into the receptacle shown in Figure 3-2. Use external
fuses for each 2000RS series instrument as 24 vdc power is distributed.
External fuse recommendation is 2 A 250 V Type T, slow blow.
CAUTION:
POWER IS APPLIED TO THE INSTRUMENT'S
CIRCUITRY AS LONG AS THE INSTRUMENT IS
CONNECTED TO THE POWER SOURCE.
3.7.2
50-P
E
IN
QUIPMENT
Figure 3-3 shows the pin layout of the Equipment Interface
connector. The arrangement is shown as seen when the viewer faces the
rear panel of the analyzer. The pin numbers for each input/output
function are given where each function is described in the paragraphs
below.

Figure 3-3: Equipment Interface Connector Pin Arrangement

Teledyne Analytical Instruments
P
OWER
I
C
NTERFACE
ONNECTOR
Installation
19