Calibration Quality; Table 7-5: Calibration Data Quality Evaluation - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Calibration Procedures
functions, slope and offset values be checked frequently to assure high quality and
accurate data from the instrument.

7.9. Calibration Quality

After completing one of the calibration procedures described above, it is important to
evaluate the analyzer's calibration SLOPE and OFFSET parameters. These values
describe the linear response curve of the analyzer. The values for these terms, both
individually and relative to each other, indicate the quality of the calibration. To perform
this quality evaluation, you will need to record the values of both test functions (Section
6.2.1 or Appendix A-3), all of which are automatically stored in the iDAS channel
CALDAT for data analysis, documentation and archival.
Make sure that these parameters are within the limits listed in Table 7-5.
Function
SLOPE
OFFS
These values should not be significantly different from the values recorded on the T-API
Final Test and Validation Data sheet that was shipped with your instrument. If they are,
refer to the troubleshooting Chapter 11.
142
Table 7-5:
Calibration Data Quality Evaluation
Minimum Value
-0.700
50.0 mV
Model 100E Instruction Manual
Optimum Value
1.000
n/a
Maximum Value
1.300
250.0 mV
045150102 Rev XB1

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents