TELEDYNE Model 9110AH NO
Table 7-8: Calibration Requirements for AutoRange or Remote Range
No matter which calibration method you choose from this chapter, all of the following
parameters must be set to properly calibrate the instrument on AutoRange or Remote range.
Parameter
Range
Slope
Offset
Converter Efficiency

7.8 Calibration Requirements for Independent Range

In Independent Range mode there is one set of calibration parameters. That is, NO
slopes and NO
x
M9110AH using independent range is the same as calibration in single range. To achieve
independent ranges, the computer scales the 3 outputs using the ranges keyed in by the
operator.

7.9 Calibration Quality

After calibration is complete, it is very important to check the QUALITY of the calibration. The
calibration of the M9110AH involves balancing several sections of electronics and software to
achieve optimum accuracy, noise, linearity and dynamic range.
The slope and offset parameters are similar to the span and zero pots on an analog instrument.
Just as in the analog instrument, if the slope or offset get outside of a certain range, the
instrument will not perform as well.
The slope value will be slightly different on the NO and NO
differences in pneumatic resistance in each pathway. If the slopes are significantly different,
there is a calibration error or a cross port leak in the switching valve. If there is a sudden
change in slopes after a calibration, that usually indicates a change in reaction cell pressure.
Check the Troubleshooting Section 9.
Analyzer Operator Manual, 01620, Rev. F
X
Low Range
Low Range is selected by pressing
CONC-LO on the front panel.
NO
and NO slopes for low range
x
are set by span calibration
NO
and NO offsets for low range
x
are set by zero calibration
CE can be set by one of the
procedures in Section 7.10.
Converter efficiency is optional.
and NO offsets, and one converter efficiency parameter. The calibration of the
High Range
High Range is selected by pressing
CONC-HI on the front panel
NO
are set by span calibration
NO
are set by zero calibration
CE can be set by one of the
procedures in Section 7.10.
Converter efficiency is optional.
channels. This is due to slight
x
7-10
and NO slopes for high range
x
and NO offsets for high range
x
and NO
x