Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series
c. Modify the initialized front-panel control settings:
Set the vertical Scale to 50 mV per division.
< 4 GHz models: Touch Vertical, select Vertical Setup, and then
touch Termination 50 Ω .
Set the horizontal Scale to 1 ns per division.
Adjust the sine-wave generator output for 4 divisions of amplitude
centered on the display.
Adjust the trigger Level to trigger at 25% (-1 division) on the sine
wave.

Table 3-13: Serial pattern data

Serial pattern data
B6DB 6DB6 DB6D B6DB
6DB6 DB6D B6DB 6DB6
DB6D B6DB 6DB6 DB6D
2. Verify that the signal path can do isolated 0 and pattern matching circuits
can do isolated 1:
a. From the button bar, touch Cursors and then the Setup button. If using
the menu bar, touch Cursors and then select Cursor Setup. Touch the
Cursor button to toggle it on and display the cursors.
b. Set the Tracking Mode to Tracking.
c. Touch the X (close) button.
d. From the button bar, touch Trig, select the A Event tab, and touch the
Select button.
e. Touch the Serial button, set Coding to NRZ, and then set the Standard
to GB Ethernet.
Trigger location
One UI before the 0
16
At the 0
16
One UI after the 0
16
3–105