Table 4-3: Maximum Bandwidth Frequency Worksheet - Tektronix MSO70000 Series Technical Reference

Specifications and performance verification
Hide thumbs Also See for MSO70000 Series:
Table of Contents

Advertisement

MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series

Table 4-3: Maximum bandwidth frequency worksheet

Model: MSO5204, DPO5204
Impedance
50 Ω
1 MΩ
Model: MSO5104, DPO5104
Impedance
50 Ω
1 MΩ
Model: MSO5054, DPO5054
50 Ω
1 MΩ
Model: MSO5034, DPO5034
50 Ω
1 MΩ
1
For DPO5204, MSO5204, DPO5104, and MSO5104 bandwidth performance verification, use 500 MHz on the
10 mV/div and 5 mV/div vertical scale.
14. Use the values of V
bw-pp
at bandwidth with the following equation:
Gain = V
/ V
bw-pp
in-pp
To pass the performance measurement test, Gain should be ≥ 0.707. Enter
Gain in the test record.
Performance Tests (MSO/DPO5000 Series)
Vertical Scale
10 mV/div — 1 V/div
5 mV/div
2 mV/div
1 mV/div
5 mV/div — 1 V/div
2 mV/div
1 mV/div
Vertical Scale
5 mV/div — 1 V/div
2 mV/div
1 mV/div
5 mV/div — 1 V/div
2 mV/div
1 mV/div
5 mV/div — 1 V/div
2 mV/div
1 mV/div
5 mV/div — 1 V/div
2 mV/div
1 mV/div
2 mV/div — 1 V/div
1 mV/div
5 mV/div — 1 V/div
2 mV/div
1 mV/div
and V
stored in the test record to calculate the Gain
in-pp
Maximum bandwidth
500 MHz
1
500 MHz
1
350 MHz
175 MHz
500 MHz
350 MHz
175 MHz
Maximum bandwidth
1
500 MHz
350 MHz
175 MHz
500 MHz
350 MHz
175 MHz
500 MHz
350 MHz
175 MHz
500 MHz
350 MHz
175 MHz
350 MHz
175 MHz
350 MHz
350 MHz
175 MHz
4–37

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSO70000 Series and is the answer not in the manual?

Questions and answers

Table of Contents