Figure 2-34: Initial test hookup

MSO70000, DSA/DPO70000/B, DPO7000 Series Specifications and Performance Verification
1. Install the test hookup and preset the instrument controls:
a. Hook Up the test-signal source: (See Figure 2-34.)
Connect the sine wave output of the sine-wave generator through a
50 Ω precision coaxial cable to Ch 1 through an adapter.
Set the sine-wave generator to output a 416 MHz sine wave.
b. Initialize the instrument: Press the Default Setup button.
c. Modify the initialized front-panel control settings:
Set the vertical Scale to 50 mV per division.
< 4 GHz models: Touch Vertical, select Vertical Setup, and then
touch Termination 50 Ω .
Set the horizontal Scale to 1 ns per division.
Adjust the sine-wave generator output for 4 divisions of amplitude
centered on the display.
Adjust the trigger Level to trigger at 25% (-1 division) on the sine
wave.

Table 2-14: Serial pattern data

Serial pattern data
B6DB 6DB6 DB6D B6DB
6DB6 DB6D B6DB 6DB6
DB6D B6DB 6DB6 DB6D
Trigger location
One UI before the 0
16
At the 0
16
One UI after the 0
16
Performance Tests
2–117