Tektronix MSO70000 Series Technical Reference page 262

Specifications and performance verification
Hide thumbs Also See for MSO70000 Series:
Table of Contents

Advertisement

Performance Tests (MSO/DPO5000 Series)
4–48
4. Record the Low and High measurements at 1 MΩ as follows:
a. Push the front-panel Wave Inspector Measure button.
b. Ensure that the Amplitude tab is selected.
c. Select the Low measurement.
d. Enter the Low measurement reading in the test record.
e. Select the High measurement.
f.
Enter the High measurement reading in the test record.
5. Record the Low and High measurements at 50 Ω as follows:
a. Push the front-panel oscilloscope Vertical Menu button.
b. Set the Termination to 50 Ω.
c. Push the front-panel Wave Inspector Measure button.
d. Ensure that the Amplitude tab is selected.
e. Select the Low measurement.
f.
Enter the Low measurement reading in the test record.
g. Select the High measurement.
h. Enter the High measurement reading in the test record.
This completes the performance verification procedure.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MSO70000 Series and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents