Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series
f.
Check against limits:
CHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings. Enter the value on the test record.
Repeat substep d, using the offset and generator settings as is listed
in the table.
CHECK that the Mean measurement readout on screen is
within the limits listed for the current vertical scale setting and
position/offset/generator settings. Enter the value on the test record.
Repeat substep d, using the negative-polarity offset and generator
settings as is listed in the table.
CHECK that the Mean measurement readout on screen is
within the limits listed for the current vertical scale setting and
position/offset/generator settings. Enter the value on the test record.
Repeat substeps c through f until all vertical scale settings, are checked
for the channel under test. (See Table 3-6.)
g. Test all channels: Repeat substeps a through f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the generator from the channel last tested.