Table 1-1: Channel input and vertical specifications, all MSO70000/C, DSA/DPO70000B/C, and DPO7000 Series
models (cont.)
Characteristic
Analog bandwidth with
TAP2500 VPI probe, typical,
DPO7254
Analog bandwidth with X10
passive probe, typical,
< 4 GHz models
Analog bandwidth selections,
< 4 GHz models
Lower frequency limit, AC coupled,
< 4 GHz models
Upper frequency limit, 250 MHz
bandwidth limited,
< 4 GHz models
Upper frequency limit, 20 MHz
bandwidth limited,
< 4 GHz models
Passband flatness, BWE, typical
MSO/DSA/DPO72004C,
MSO/DSA/DPO71604C,
MSO/DSA/DPO71254C
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series
Specifications (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
Description
Full bandwidth, operating ambient of ≤ 30 °C (86 °F), derated by 1% for each °C above
30 °C (86 °F)
SCALE range
≥ 100 mV/div
50 mV/div to 99.5 mV/div
20 mV/div to 49.8 mV/div
10 mV/div to 19.9 mV/div
Full bandwidth, operating ambient of ≤ 30 °C (86 °F), derated by 1% for each °C above
30 °C (86 °F)
SCALE range
5 mV/div to 10 V/div
2 mV/div to 4.98 mV/div
1 mV/div to 1.99 mV/div
20 MHz, 250 MHz, and Full
10 Hz when 1 MΩ, AC coupled. The limit is reduced by a factor of 10 when 10X, passive
probes are used.
250 MHz
20 MHz
Use the Temperature variation table to determine the amount of performance derating above
the temperature limit.
Step settings
100 mVFS, 200 mVFS, 500 m VFS, 1 VFS, and
2.5 VFS
± 0.5 dB from DC to 50% of nominal bandwidth.
± 1.5 dB from 50% to 80% of nominal bandwidth.
Non-step settings
± 1.0 dB from DC to 50% of nominal bandwidth
± 2.0 dB from 50% to 80% of nominal bandwidth.
Bandwidth
DC to 2.5 GHz
DC to 2.0 GHz
DC to 500 MHz
DC to 200 MHz
Bandwidth
DC to 500 MHz
DC to 300 MHz
DC to 175 MHz
Temperature
constraint
At 25 °C
1–9