Mixed signal oscilloscopes, digital serial analyzers, digital phosphor oscilloscopes, digital phosphor oscilloscopes declassification and security (26 pages)
Page 1
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, and DPO5000/B Series Oscilloscopes Specifications and Performance Verification Technical Reference *P077006314* 077-0063-14...
Page 3
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, and DPO5000/B Series Oscilloscopes Specifications and Performance Verification Technical Reference Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so.
Page 4
Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material.
Page 5
Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
General safety summary ........................ Terms in the manual ........................Terms on the product ........................Symbols on the product ......................... Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels ..................... Horizontal and acquisition system ....................Trigger specifications ........................Serial trigger specifications ......................
Page 8
Check analog bandwidth ......................Check random noise sample acquisition mode ................Check sample rate and delay time accuracy ................Check delta time measurement accuracy ..................Check digital threshold accuracy (MSO5000/B only) ............... Check trigger out ........................MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Before use, always check the product with a known source to be sure it is operating correctly. This product is not intended for detection of hazardous voltages. Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 10
Use only insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product. Observe all terminal ratings. To avoid fire or shock hazard, observe all rating and markings on the product.
Page 11
WARNING. The product is heavy. To reduce the risk of personal injury or damage to the device get help when lifting or carrying the product. Use only the Tektronix rackmount hardware specified for this product. Probes and test leads Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
(This symbol may also be used to refer the user to ratings in the manual.) The following symbols may appear on the product: MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) To meet specifications, two conditions must first be met: ■ The instrument must have been calibrated in an ambient temperature between 18 °C and 28 °C (64 °F and 82 °F). ■ The instrument must be operating within the environmental limits. (See page 1-78.)
Page 15
DPO70K and DPO7K series only Category II, III or IV circuits. Input termination voltage (VTerm) range >20 GHz models ≤1.2 VFS settings: -3.5 V to +3.5 V >1.2 VFS settings: Input termination voltage (VTerm) range ≤20 GHz models MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 16
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) Input VSWR >20 GHz models, Measured with a TekConnect TCA-292D adaptor and a network analyzer. typical ≤1.2 VFS settings: 0 – 17 GHz: 1.4:1 17 – 20 GHz: 1.6:1 20 –...
Page 17
±(DC Gain Accuracy × | reading -(net offset) | + offset accuracy + 0.1 division) ( ≥16 averages) Delta voltage measurement ±(DC Gain Accuracy × | reading | + 0.05 division) between any two averages of ≥16 waveforms acquired with the same setup and ambient conditions MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 18
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) Sample acquisition mode, ±(DC Gain Accuracy × | reading -(net offset) | + offset accuracy + 0.15 division + 0.6 mV) typical Delta voltage measurement ±(DC Gain Accuracy × | reading | + 0.15 division + 1.2 mV)
Page 19
±1.90 V 140 mV/div ±1.80 V 160 mV/div ±1.70 V 200 mV/div ±1.50 V 248 mV/div ±1.260 V 250 mV/div ±1.250 V 300 mV/div ±1.00 V 400 mV/div ±0.50 V 500 mV/div ±0.00 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 20
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) <4 GHz models, 50 Ω coupling SCALE range Offset range 1 mV/div to 50 mV/div ±1.0 V 50.5 mV/div to 99.5 mV/div ±1.5 V - 10 divisions 100 mV/div to 500 mV/div ±10 V...
Page 21
2.5 V. Enhanced bandwidth of 4 GHz, 6 GHz, and 8 GHz is only available at 25 GS/s and 50 GS/s. Use the Temperature variation table to determine the amount of performance derating above the temperature limit. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 22
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) The MSO/DPO72004C and MSO/DPO71604C instrument bandwidths with BWE disabled are typical. MSO/DPO72004C BW Settings Bandwidth Temp constraint 20 GHz BWE DC to >20 GHz <30 °C 20 mV, 50 mV, 100 mV, and...
Page 23
≥ 100 mV/div DC to 2.5 GHz 50 mV/div to 99.5 mV/div DC to 2.0 GHz 20 mV/div to 49.8 mV/div DC to 500 MHz 10 mV/div to 19.9 mV/div DC to 200 MHz MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 24
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) Analog bandwidth with X10 Full bandwidth, operating ambient of ≤ 30 °C (86 °F), derated by 1% for each °C above 30 °C passive probe, typical, <4 GHz (86 °F)
Page 28
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) Pulse response peak detect or This instrument uses analog peak detection for pulse capture in Peak Detect or Envelope mode at envelope mode <4 GHz models sample rates of 125 MS per second and slower. At faster sample rates the instrument uses conventional sampling.
Page 30
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Vertical system analog channels (cont.) DSP on, full bandwidth Nine division sine wave input at the indicated frequency, sampled at 500 mV vertical sensitivity and maximum sample rate MSO/DPO73304DX MSO/DPO72504DX MSO/DPO72304DX Input 50 GS/s...
Page 59
50 Ω, DC coupling and scale set at or above 10 mV/div. Internal probe power <4 GHz 50 W maximum models Probe power per probe, <4 GHz Voltage Maximum Amperage Voltage Tolerance models 50 mA ± 5% 12 V ±10% MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Horizontal and acquisition system Real-time sample rate range Number of channels Sample rate, maximum acquired (Standard) (Option 2SR) MSO/DPO70804C All channels Up to 25 GS/s MSO/DPO70604C MSO/DPO70404C MSO/DPO73304DX All channels Up to 50 GS/s...
Page 61
125,000,000 points (3 or 4 channels) 20 GS/s or less 250,000,000 points (2 channels only) 40 GS/s or less 500,000,000 points (1 channel only) Maximum record length, HiRes Half the record length of sample mode mode MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 62
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Horizontal and acquisition system (cont.) Seconds division range ≥4 GHz models, Fastest sweep speed is 20 ps per division <4 GHz models DPO7354 Record Length DPO7254 Horizontal 10 K 100 K 10 M 100 M...
Page 63
1.5 V/ns. This is most easily achieved either with a square wave meeting that slew rate or using a 100 MHz sine wave (≈2 Vp-p). This is required for the best possible performance in synchronous sampling applications. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 64
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Horizontal and acquisition system (cont.) <4 GHz models Total: <1 ps rms for record durations less than 10 μs. <2.5 ps rms for record durations less than 30 ms. <65 parts/trillion for record durations less than 10 s.
Page 65
(the amount of time from the first single-shot measurement to the final single-shot measurement). The observation interval may not exceed 1 year. Test at maximum bandwidth. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 66
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Horizontal and acquisition system (cont.) Test at 62.5 mVFS, 50 mVFS, 100 mVFS. MSO/DPO73304DX at 6.25 mV/div, <10 ns measurement interval, 19.49 GHz, amplitude = 80% FS, 0.803 ps rms, 100 GS/s, BWE on 8.03 ps...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Trigger specifications Trigger jitter DC coupled A edge random holdoff, typical ≥4 GHz models 100 fs using enhanced trigger placement. 1 ps rms for low frequency, fast rise time signal, A edge, time holdoff = 30 μs <4 GHz models...
Page 71
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Edge trigger sensitivity not DC All sources, positive or negative edge, for vertical scale settings ≥10 mV/div and ≤1 V/div coupled, typical ≥4 GHz models Trigger Coupling Sensitivity NOISE REJ 15%FS from DC to 50 MHz...
Page 72
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Trigger specifications (cont.) Trigger level or threshold Edge trigger, DC coupled accuracy, typical ≥4 GHz models Trigger Source Accuracy Channel 1, 2, 3, or 4 ± [(2% | trigger level - net offset | ) + (3.5% FS)
Page 73
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Width and glitch trigger sensitivity, typical ≥4 GHz models 15% x FS from DC to 1 GHz, not violating the minimum timing requirements for each type (for vertical settings from 10 mV/div to 1 V/div at the input connector) <4 GHz models...
Page 74
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Trigger specifications (cont.) Pattern and state trigger sensitivity 15% x FS from DC to 1 GHz, not violating the minimum timing requirements for each type (for DC coupled, typical, ≥4 GHz vertical settings from 10 mV/div to 1 V/div at input connector)
Page 75
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Window trigger minimum timing requirements ≥4 GHz models Minimum Minimum Setup time Hold time pulse width rearm time Window enters 150 ps 500 ps N.A. N.A. Window exits 500 ps 150 ps N.A.
Page 76
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Trigger specifications (cont.) Setup-hold time violation trigger, minimum clock pulse widths, typical ≥4 GHz models Minimum time from active clock edge to Minimum time from inactive clock edge to inactive edge active edge Hold times ≥ 0: Hold time + 550 ps 500 ps Hold times <0: Hold time + 1.6 ns...
Trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Variable A event trigger holdoff 250 ns to 12 s + random holdoff range Lowest frequency for successful 50 Hz set level to 50%, typical Trigger jitter (RMS) ≥4 GHz models 100 fs using enhanced Trigger placement.
Page 78
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Serial trigger specifications (cont.) High speed serial trigger, serial word recognizer position accuracy ≥4 GHz models ± 200 ps <4 GHz models Acquisition mode Sample, Average ± (1 waveform interval + 200 ps) Peak Detect, Envelope ±...
Page 79
Serial trigger specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Clock recovery jitter, typical ≥4 GHz models < 0.25% bit period + 2 ps rms for PRBS data patterns with 50% transition density. < 0.25% bit period + 1.5 ps rms for repeating 0011 data patterns.
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Digital acquisition specifications Minimum signal amplitude 300 mV p-p; logic threshold centered in dynamic range of signal; f ≤1.5 GHz for P6780 f ≤350 MHz for P6717A Minimum detectable pulse width <400 ps (mainframe and logic probe),...
Page 81
12.5 GS/s or less 20,000,000 points (all channels) Option 5XL installed 12.5 GS/s or less 50,000,000 points (all channels) Option 10XL installed 12.5 GS/s or less 125,000,000 points (all channels) Option 20XL installed 12.5 GS/s or less 250,000,000 points (all channels) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Input-output port specifications Auxiliary trigger input characteristics and range ≥4 GHz models 50 Ω, ±4 V (DC plus peak AC) MSO/DPO70000DX: 50 Ω, ±5 V (DC plus peak AC) <4 GHz models 50 Ω, ± 5%; maximum input voltage ± 5 V (DC + peak AC) 1 MΩ, ±5%, 150 V CAT1.
Page 83
5 ps RMS trigger External reference input frequency, Run SPC whenever the external reference is more than 0.1% (1000 ppm) different than the typical nominal reference or the reference at which SPC was last run. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 84
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Input-output port specifications (cont.) The timebase changes in correspondence to the fluctuations in the external reference. ≥4 GHz models 10 MHz, 100 MHz The instrument scans for either 10 MHz or 100 MHz. <4 GHz models...
2.5-inch SATA hard disk, drive capacity is ≥300 GB. DX models use a ≥300 GB solid state drive. C models use a ≥500 GB hard disk drive. ≤20 GHz models 2.5-inch SATA hard disk, drive capacity is ≥160 GB. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Power source specification Power consumption ≥4 GHz models <1200 VA <4 GHz models 550 Watts maximum Source voltage and frequency ≥4 GHz models 100 to 240 V , 50 Hz to 60 Hz, 115 V ± 10%, 400 Hz CAT II <4 GHz models...
Page 87
Mechanical specifications (cont.) Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Dimensions <4 GHz models, Benchtop With front cover and feet Without front cover and feet configuration 295.4 mm (11.6 in) height 278 mm (10.95 in) height 468.6 mm (18.4 in) width 450.8 mm (17.75 in) width...
Specifications (MSO70000C/DX, DPO70000C/DX, and DPO7000C series) Environmental specifications Temperature <4 GHz models Operating +5 °C to +45 °C (41 °F to +113 °F), with noncondensing conditions Nonoperating -40 °C to +71 °C (-40 °F to +160 °F), with 15 °C/hour maximum gradient, without disk media installed in disk drives Temperature ≥4 GHz models...
VSWR ≤1.5:1 from DC to 1 GHz, typical DPO5104/B MSO5054/B, VSWR ≤1.5:1 from DC to 500 MHz, typical DPO5054B MSO5034/B, VSWR ≤1.5:1 from DC to 350 MHz, typical DPO5034/B Input capacitance 1MΩ DC coupled 13 pF ± 2 pF MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 90
For 50 Ω, 1 MΩ, TPP0500, and TPP1000 path: ±1.5%, derated at 0.100%/°C above 30 °C ±2.0%, derated at 0.100%/°C above 30 °C, 1 mV/Div setting, typical ±3.0% variable gain, derated at 0.100%/°C above 30 °C MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 91
The delta volts (difference voltage) accuracy specification applies directly to the following measurements: Positive Overshoot, Negative Overshoot, Peak-to-Peak, and Amplitude. DC voltage measurement accuracy, sample acquisition mode, typical MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 92
1 mV/div — 1.99 mV/div DC to 175 MHz MSO5034/B, 5 mV/div — 10 V/div DC to 350 MHz DPO5034/B 2 mV/div — 4.98 mV/div DC to 350 MHz 1 mV/div — 1.99 mV/div DC to 175 MHz MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 97
15 W maximum internal probe power (12 W maximum with firmware versions <6.4.0) Provision for 50 W probe power with 119-7465-00 TekVPI External Power Supply Probe power per channel Voltage Max Amperage Voltage Tolerance 50 mA (250 mW) ±5% 12 V 2 A (24 W) ±10% MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
The record length ranges for a standard instrument are the same as for the analog channel specification table (above). The record length ranges for an instrument with an option are the same as for the analog channel specifications, up to 40M. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 99
N =input-referred noise (VRMS) TBA = time base accuracy (5 ppm) tp = delta-time measurement duration (seconds) RD = (record length)/(sample rate) tsr = 1/(sample rate) assume edge shape that results from Gaussian filter response MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 101
Triggers from negative sync composite video, field 1 or field 2 for interlaced systems, on any field, field rates specific line, or any line for interlaced or non-interlaced systems. Supported systems include NTSC, PAL, and SECAM. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 102
1 ns 1.0 ms Setup + Hold time 0.5 ns 2.0 ms Input coupling on clock and data channels must be the same. For Setup time, positive numbers mean a data transition before the clock. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 103
Trigger level ranges Source Range Any input channel ±8 divisions from center of screen ±8 divisions from 0 V when vertical LF reject trigger coupling is selected Aux In (External) ±8 V Line Not applicable MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 104
Data Rates Supported: HS to 480 Mbs, Full to 12 Mbs, Low to 1.5 Mbs Trigger On: Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data Packet, Handshake Packet, Special Packet, Error MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
±42 V peak at the P6616 input (not at the instrument input) Probe input voltages beyond this limit could permanently damage the instrument and the P6616 probe. Channel-to-channel skew 200 ps Digital channel to digital channel only. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
You can set the Auxiliary output to Trigger Out or Reference Clock Out. out or reference clock out Reference Clock Out: Outputs the 10 MHz oscilloscope reference clock. Trigger Out: A HIGH to LOW transition indicates that the trigger occurred. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 107
Rin = 1.5 kΩ ±20% in parallel with 15 pF ±5 pF at 10 MHz impedance Keyboard and mouse ports PS-2 compatible, instrument must be powered down to make connection External audio ports External audio jacks for microphone input and line output MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Source frequency (85 to 264 V) 45 Hz to 66 Hz (100 V to 132 V) 360 Hz to 440 Hz Fuse rating T6.3AH, 250 V The fuse cannot be replaced by the user. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
High: 40 °C to 60 °C (104 °F to 140 °F), 5% to 60% relative humidity Low: 0 °C to 40 °C (32 °F to 104 °F), 5% to 90% relative humidity Altitude Operating: 3,000 m (9,843 ft) Nonoperating: 12,000 m (39,370 ft) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
2 in (50.8 mm) Cooling Fan-forced air circulation with no air filter. Construction material Chassis parts are constructed of aluminum alloy, front panel is constructed of plastic laminate, circuit boards are constructed of glass laminate. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Performance tests (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series) on page 115 Advantages. These procedures add direct checking of the warranted specifications that are marked with the symbol. These procedures require specific test equipment. Equipment required on page 116 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
In steps and substeps, the lead-in statement in italics instructs you what to do, while the instructions that follow tell you how to do it, as in the example step below: Initialize the instrument: Push the front-panel Default Setup button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 113
The procedures assume you have connected a mouse to the instrument so you can click on the screen controls. If you have not connected a mouse, you can use the touch screen to operate the screen controls. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
This displays the instrument calibration control window. If required because the instrument is in service mode, select the Signal Path button under Calibration Area. Click the Run SPC (Calibrate on some instruments) button to start the routine. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
When you are instructed to press a front-panel or screen button, the button may already be selected (its label will be highlighted). If this is the case, it is not necessary to press the button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 116
3. Turn off all channels: If any of the front-panel channel buttons are lighted, push those buttons to turn off the displayed channels as shown in the following figure. Figure 3: Channel button location MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 117
(See Table 4: Vertical settings on page 105.) 7. Verify that the channel acquires in all acquisition modes: Pull down the Horiz/Acq menu to select Horizontal/Acquisition Setup..Click the MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 118
Waveform Database mode displays an actively acquiring waveform on- screen with the noise displayed. 8. Test all channels: Repeat steps 2 through 7 until all four input channels are verified. 9. Remove the test hookup: Disconnect the equipment from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 119
(more horizontal divisions per waveform period), counterclockwise rotation contracts it, and returning the horizontal scale to 200 μs/div returns the period to about five divisions. ■ The horizontal Position knob positions the signal left and right on-screen when rotated. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 120
Verify that the display switches quickly between two different points in time (the rising edge shifts horizontally on the display). 8. Remove the test hookup: Disconnect the test hookup from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 121
Set up the delayed trigger: Pull down the Trig menu and select A – B Trigger Sequence..This displays the A →B Sequence tab of the trigger setup control window. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 122
■ Verify that the trigger Ready indicator on the front panel flashes about once every second as the waveform is updated on-screen. 7. Remove the test hookup: Disconnect the test hookup from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 123
7. Change the settings again: Set the horizontal SCALE to 200μs/div. 8. Verify the file system works: a. Pull down the File menu to select Recall. This displays the instrument Recall control window. b. Click Recall What > Setup. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 124
SCALE is again 1 ms and the waveform shows ten cycles just as it did when you saved the setup. 9. Remove the test hookup: Disconnect the test hookup from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 125
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series) Verify the digital channels (MSO70000C/DX series only). Equipment required Prerequisites MSO70000C/DX Series models: One precision 50 Ω None coaxial cable (item 4) MSO70000C/DX Series models: One probe tip adapter (item 30) MSO70000C/DX Series models: One SMA adapter...
Page 126
12. Verify that the both the iCapture and the digital channel are triggered. 13. Rotate the Vertical SCALE knob counter clockwise. 14. Verify that the displayed amplitude of channel 1 is reduced. 15. Rotate the Vertical SCALE knob clockwise, back to its original setting. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
These procedures operate the instrument near its maximum limits. Using settings similar to, but not the settings called for by the procedures, could damage the input circuitry of the instrument. See the Maximum input voltage specification for the maximum nondestructive input voltage level. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
2. Attenuator, 5X Ratio: 5X; impedance 50 Ω; BNC ≤ 2 GHz Tektronix part Signal attenuation, bandwidth, connectors: female input, number 011-0060-02 trigger sensitivity male output SMA ≤ 18 GHz Tektronix part number 015-1002-01 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 129
10 MHz and 100 MHz, 1 s Advantest R5360 Checking long-term sample gate rate and delay time accuracy Fluke 9500B/1100, 9500B/3200, or 9500B/2200 and an output head (9510 or 9530) appropriate for the bandwidth of the instrument being tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 130
>3 GHz: Frequency range: DC to 40 GHz. Tracking: <2.0% 12. Cable 2.92 mm male-to-female Gore PhaseFlex cable Checking analog bandwidth EL0CQ0CP0360 (40 GHz) 13. Adapter K male-to-male DC to 40 GHz Anritsu K220B Checking analog bandwidth MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 131
Checking delta time SMA connectors 174-1427-00 measurement accuracy, probe compensation out, trigger sensitivity For Delta Time Measurement Accuracy, use a Fluke 9500B or a pulse generator with the rise time specified in the procedure. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 132
29. Mouse or keyboard Tektronix part numbers: Used to input test selections 119-7054-xx (mouse) 119-7083-xx (keyboard) 30. Probe tip adapter BNC female to square pin Tektronix part number Checking digital channels adapter, 50 Ω 067-1734-00 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series) 31. Logic probe P6717 or P6780 Logic Probe Tektronix part P6717 or Checking digital channels P6780 32. Adapter BNC “T”, male to 2 females Tektronix part number Checking DC gain and offset 103-0030-xx accuracy 33.
Page 134
__________ __________ -2.4463 V -1.5 V termination -2.5 V offset -2.5312 V __________ __________ -2.4988 V -1.5 V offset -1.5405 V __________ __________ -1.5195 V -0.5 V offset -538.5 mV __________ __________ -521.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 135
__________ __________ -2.5257 V -1.5 V termination -2.5 V offset -2.6745 V __________ __________ -2.6255 V -1.5 V offset -1.6685 V __________ __________ -1.6315 V -0.5 V offset -666.5 mV __________ __________ -633.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 136
Ch1 200 mV Vert scale setting, 0 Div position setting 0 V termination –1 V offset -456.5 mV __________ __________ -343.5 mV 0 V offset 547.5 mV __________ __________ 652.5 mV +1 V offset 1.5435 V __________ __________ 1.6565 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 137
__________ __________ -2.4463 V -1.5 V termination -2.5 V offset -2.5312 V __________ __________ -2.4988 V -1.5 V offset -1.5405 V __________ __________ -1.5195 V -0.5 V offset -538.5 mV __________ __________ -521.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 138
__________ __________ -2.5257 V -1.5 V termination -2.5 V offset -2.6745 V __________ __________ -2.6255 V -1.5 V offset -1.6685 V __________ __________ -1.6315 V -0.5 V offset -666.5 mV __________ __________ -633.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 139
Ch2 200 mV Vert scale setting, 0 Div position setting 0 V termination –1 V offset -456.5 mV __________ __________ -343.5 mV 0 V offset 547.5 mV __________ __________ 652.5 mV +1 V offset 1.5435 V __________ __________ 1.6565 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 140
__________ __________ -2.4463 V -1.5 V termination -2.5 V offset -2.5312 V __________ __________ -2.4988 V -1.5 V offset -1.5405 V __________ __________ -1.5195 V -0.5 V offset -538.5 mV __________ __________ -521.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 141
__________ __________ -2.5257 V -1.5 V termination -2.5 V offset -2.6745 V __________ __________ -2.6255 V -1.5 V offset -1.6685 V __________ __________ -1.6315 V -0.5 V offset -666.5 mV __________ __________ -633.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 142
Ch3 200 mV Vert scale setting, 0 Div position setting 0 V termination –1 V offset -456.5 mV __________ __________ -343.5 mV 0 V offset 547.5 mV __________ __________ 652.5 mV +1 V offset 1.5435 V __________ __________ 1.6565 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 143
__________ __________ -2.4463 V -1.5 V termination -2.5 V offset -2.5312 V __________ __________ -2.4988 V -1.5 V offset -1.5405 V __________ __________ -1.5195 V -0.5 V offset -538.5 mV __________ __________ -521.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 144
__________ __________ -2.5257 V -1.5 V termination -2.5 V offset -2.6745 V __________ __________ -2.6255 V -1.5 V offset -1.6685 V __________ __________ -1.6315 V -0.5 V offset -666.5 mV __________ __________ -633.5 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 145
Ch4 140 mV Vert scale setting, 0 Div position setting 0 V termination –1 V offset -624.5 mV __________ __________ -535.5 mV 0 V offset 379.5 mV __________ __________ 460.5 mV +1 V offset 1.3755 V __________ __________ 1.4645 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 146
- 535.25 mV __________ __________ - 524.75 mV +5 Div position setting, -0.45 V offset Ch2 20 mV Vert scale setting, + 552.75 mV __________ __________ + 567.25 mV -5 Div position setting, +0.4 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 147
+ 3.394 V __________ __________ + 3.606 V -5 Div position setting, 0 V offset Ch3 500 mV Vert scale setting, - 3.606 V __________ __________ - 3.394 V +5 Div position setting, 0 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 148
+ 574.19 mV __________ __________ + 585.81 mV -5 Div position setting, +0.5 V offset Ch1 10 mV Vert scale setting, - 585.81 mV __________ __________ - 574.19 mV +5 Div position setting, -0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 149
- 543.74 mV __________ __________ - 536.26 mV +5 Div position setting, -0.5 V offset Ch2 10 mV Vert scale setting, + 574.19 mV __________ __________ + 585.81 mV -5 Div position setting, +0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 150
+ 536.26 mV __________ __________ + 543.74 mV -5 Div position setting, +0.5 V offset Ch3 5 mV Vert scale setting, - 543.74 mV __________ __________ - 536.26 mV +5 Div position setting, -0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 151
- 519.03 mV __________ __________ - 512.97 mV +5 Div position setting, -0.5 V offset Ch4 5 mV Vert scale setting, + 536.26 mV __________ __________ + 543.74 mV -5 Div position setting, +0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 152
- 509.76 mV __________ __________ - 504.24 mV +5 Div position setting, -0.5 V offset Ch1 2 mV Vert scale setting, + 512.97 mV __________ __________ + 519.03 mV -5 Div position setting, +0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 153
+ 504.24 mV __________ __________ + 509.76 mV -5 Div position setting, +0.5 V offset Ch2 1 mV Vert scale setting, - 509.76 mV __________ __________ - 504.24 mV +5 Div position setting, -0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 154
- 99.610 V __________ __________ - 98.390 V 0 Div position setting, -96.0 V offset Ch3 1 mV Vert scale setting, + 504.24 mV __________ __________ + 509.76 mV -5 Div position setting, +0.5 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 155
+ 98.390 V __________ __________ + 99.610 V 0 Div position setting, +96.0 V offset Ch3 1.0 V Vert scale setting, - 99.610 V __________ __________ - 98.390 V 0 Div position setting, -96.0 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 156
+ 4.345 V __________ __________ + 4.655 V -2 Div position setting, +5.0 V offset Ch4 500 mV Vert scale setting, - 4.655 V __________ __________ - 4.345 V +2 Div position setting, -5.0 V offset MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 157
__________ + 61.2 mV 0 Div position setting, -1.5 V offset + 58.8 mV __________ __________ + 61.2 mV 0 Div position setting, -0.5 V offset + 58.8 mV __________ __________ + 61.2 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 158
__________ + 306 mV 0 Div position setting, -1.5 V offset + 294 mV __________ __________ + 306 mV 0 Div position setting, -0.5 V offset + 294 mV __________ __________ + 306 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 159
0 Div position setting, -0.9 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, 0 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, +0.9 V offset 1176 mV __________ __________ 1224 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 160
__________ + 61.2 mV 0 Div position setting, -1.5 V offset + 58.8 mV __________ __________ + 61.2 mV 0 Div position setting, -0.5 V offset + 58.8 mV __________ __________ + 61.2 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 161
__________ + 306 mV 0 Div position setting, -1.5 V offset + 294 mV __________ __________ + 306 mV 0 Div position setting, -0.5 V offset + 294 mV __________ __________ + 306 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 162
0 Div position setting, -0.9 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, 0 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, +0.9 V offset 1176 mV __________ __________ 1224 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 163
__________ + 61.2 mV 0 Div position setting, -1.5 V offset + 58.8 mV __________ __________ + 61.2 mV 0 Div position setting, -0.5 V offset + 58.8 mV __________ __________ + 61.2 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 164
__________ + 306 mV 0 Div position setting, -1.5 V offset + 294 mV __________ __________ + 306 mV 0 Div position setting, -0.5 V offset + 294 mV __________ __________ + 306 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 165
0 Div position setting, -0.9 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, 0 V offset 1176 mV __________ __________ 1224 mV 0 Div position setting, +0.9 V offset 1176 mV __________ __________ 1224 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 166
__________ + 61.2 mV 0 Div position setting, -1.5 V offset + 58.8 mV __________ __________ + 61.2 mV 0 Div position setting, -0.5 V offset + 58.8 mV __________ __________ + 61.2 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 167
__________ + 306 mV 0 Div position setting, -1.5 V offset + 294 mV __________ __________ + 306 mV 0 Div position setting, -0.5 V offset + 294 mV __________ __________ + 306 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 168
0 Div position setting, -1 V offset 823.2 mV __________ __________ 856.8 mV 0 Div position setting, 0 V offset 823.2 mV __________ __________ 856.8 mV 0 Div position setting, +1 V offset 823.2 mV __________ __________ 856.8 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 169
__________ + 155.04 mV -5 Div position setting, +0.4 V offset + 148.96 mV __________ __________ + 155.04 mV +5 Div position setting, -0.4 V offset + 148.96 mV __________ __________ + 155.04 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 170
__________ + 77.52 mV -5 Div position setting, +0.45 V offset + 74.48 mV __________ __________ + 77.52 mV +5 Div position setting, -0.45 V offset + 74.48 mV __________ __________ + 77.52 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 171
__________ + 1.55 V -5 Div position setting, +1.25 V offset + 1.849 V __________ __________ + 1.55 V +5 Div position setting, -1.25 V offset + 1.49 V __________ __________ + 1.55 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 172
__________ + 775.20 mV -5 Div position setting, +2.0 V offset + 744.80 mV __________ __________ + 775.20 mV +5 Div position setting, -2.0 V offset + 744.80 mV __________ __________ + 775.20 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 173
0 V offset, 0 V termination –3.5 mV __________ __________ +3.5 mV +3.4 V offset, +3.5 V termination +3.3827 V __________ __________ +3.4173 V -3.4 V offset, -3.5 V termination –3.4173 V __________ __________ –3.3827 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 174
0 V offset, 0 V termination –7.5 mV __________ __________ +7.5 mV +3.4 V offset, +3.5 V termination +3.3787 V __________ __________ +3.4213 V -3.4 V offset, -3.5 V termination –3.4213 V __________ __________ –3.3787 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 175
0 V offset, 0 V termination –14.5 mV __________ __________ +14.5 mV +3.4 V offset, +3.5 V termination +3.3717 V __________ __________ +3.4283 V -3.4 V offset, -3.5 V termination –3.4283 V __________ __________ –3.3717 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 176
-6 V offset, 0 V termination –6.0525 V __________ __________ –5.9475 V 0 V offset, 0 V termination –16.5 mV __________ __________ +16.5 mV +6 V offset, 0 V termination +5.9475 V __________ __________ +6.0525 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 177
Ch2 200 mV Vert scale setting, -5 Div position setting, +1.5 V offset + 2.464 V __________ __________ + 2.536 V 5 Div position setting, -1.5 V offset - 2.536 V __________ __________ - 2.464 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 178
Ch4 500 mV Vert scale setting, -5 Div position setting, 0 V offset + 2.434 V __________ __________ + 2.566 V 5 Div position setting, 0 V offset - 2.566 V __________ __________ - 2.434 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Signal acquisition system These procedures check those characteristics that relate to the signal-acquisition system and are listed as checked under Warranted Characteristics in checks Specifications. See Equipment required on page 116 for test equipment specifications. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 189
Set the Horizontal Scale to 1 ms/div. ■ Set the Record Length to 5000. ■ Set the Acquisition Mode to Hi Res. ■ Set the Trigger Source to Line. Figure 8: DC voltage accuracy test hookup, >20 GHz models MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 190
999.45 mV to 1.01856 V -3.5 V -3.4 V -3.419 V -3.4363 V to -3.4012 V -3.5 V -2.4 V -2.419 V -2.4343 V to -2.4032 V Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 191
+0.5 V 650 mV 633.5 mV to 666.5 mV 150 mV 137.5 mV to 162.5 mV –1.0 V -850 mV -868.5 mV to -831.5 mV Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 192
379.5 mV to 460.5 mV +1.0 V 1.42 V 1.3755 V to 1.4645 V 200 mV –1.0 V -400 mV -456.5 mV to -343.5 mV Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 193
Measure the test signal: Press the X (close) button. Read the measurement results at the measurement statistics Mean measurement readout. See the following figure. Figure 9: Measurement of DC accuracy Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 194
Test all channels: Repeat substeps 2.a through 2.f for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the equipment from the generator output and the input connector of the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 195
Do the following substeps - test Ch 1 first, skipping substep 2.a of this step since Ch 1 is already selected from step 1. a. Select an unchecked channel: ■ From the button bar, touch Measure and then Clear All to remove the previous measurement. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 196
-535.25 mV to -524.75 mV 20 mV +0.4 V +560 mV +552.75 mV to +567.25 mV -0.4 V -560 mV -567.25 mV to -552.75 mV Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 197
+98.390 V to +99.610 V -96.0 V -99.0 V -99.610 V to -98.390 V d. Display the test signal: ■ From the button bar touch Vertical and touch Position. Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 198
■ CHECK that the Mean measurement readout on screen is within the limits listed for the current vertical scale setting and position/offset/ generator settings. Enter the value on test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 199
1 MΩ. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the equipment from the generator output and the input connector of the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 200
From the button bar, touch Horiz/Acq and select the Acquisition tab. ■ Set the Horizontal Scale to 1 ms/div. ■ Set the Record Length to 5000. ■ Set the Acquisition Mode to Hi Res. ■ Set the Trigger Source to Line. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 201
0.5 V 519 mV +37.24 mV to +38.76 mV 481 mV 3.4 V 3.4 V 3.419 V +37.24 mV to +38.76 mV 3.381 V 2.4 V 2.419 V +37.24 mV to +38.76 mV 2.381 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 202
-1.5 V -2.5 V -2.47 V +58.8 mV to +61.2 mV -2.53 V -1.5 V -1.47 V +58.8 mV to +61.2 mV -1.53 V -0.5 V -470 mV +58.8 mV to +61.2 mV -530 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 203
-560 mV 50 mV 150 mV +294 mV to +306 mV -150 mV +1 V 1.15 V +294 mV to +306 mV 850 mV -1 V -850 mV +294 mV to +306 mV -1.15 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 204
0.5 V 800 mV +588 mV to +612 mV 200 mV 3.4 V 3.4 V 3.70 V +588 mV to +612 mV 3.10 V 2.4 V 2.70 V +588 mV to +612 mV 2.10 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 205
-2.4 V -2.04 V +705.6 mV to +734.4 mV -2.76 V -1.5 V -1.5 V -1.14 V +705.6 mV to +734.4 mV -1.86 V -0.5 V -140 mV +705.6 mV to +734.4 mV -860 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 206
0.5 V 519 mV +37.24 mV to +38.76 mV 481 mV 3.4 V 3.4 V 3.419 V +37.24 mV to +38.76 mV 3.381 V 2.4 V 2.419 V +37.24 mV to +38.76 mV 2.381 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 207
-1.5 V -2.5 V -2.47 V +58.8 mV to +61.2 mV -2.53 V -1.5 V -1.47 V +58.8 mV to +61.2 mV -1.53 V -0.5 V -470 mV +58.8 mV to +61.2 mV -530 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 208
-560 mV 50 mV 150 mV +294 mV to +306 mV -150 mV +1 V 1.15 V +294 mV to +306 mV 850 mV -1 V -850 mV +294 mV to +306 mV -1.15 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 209
0.5 V 800 mV +588 mV to +612 mV 200 mV 3.4 V 3.4 V 3.70 V +588 mV to +612 mV 3.10 V 2.4 V 2.70 V +588 mV to +612 mV 2.10 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 210
-2.4 V -2.04 V +705.6 mV to +734.4 mV -2.76 V -1.5 V -1.5 V -1.14 V +705.6 mV to +734.4 mV -1.86 V -0.5 V -140 mV +705.6 mV to +734.4 mV -860 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 211
0.5 V 519 mV +37.24 mV to +38.76 mV 481 mV 3.4 V 3.4 V 3.419 V +37.24 mV to +38.76 mV 3.381 V 2.4 V 2.419 V +37.24 mV to +38.76 mV 2.381 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 212
-1.5 V -2.5 V -2.47 V +58.8 mV to +61.2 mV -2.53 V -1.5 V -1.47 V +58.8 mV to +61.2 mV -1.53 V -0.5 V -470 mV +58.8 mV to +61.2 mV -530 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 213
-560 mV 50 mV 150 mV +294 mV to +306 mV -150 mV +1 V 1.15 V +294 mV to +306 mV 850 mV -1 V -850 mV +294 mV to +306 mV -1.15 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 214
0.5 V 800 mV +588 mV to +612 mV 200 mV 3.4 V 3.4 V 3.70 V +588 mV to +612 mV 3.10 V 2.4 V 2.70 V +588 mV to +612 mV 2.10 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 215
-2.4 V -2.04 V +705.6 mV to +734.4 mV -2.76 V -1.5 V -1.5 V -1.14 V +705.6 mV to +734.4 mV -1.86 V -0.5 V -140 mV +705.6 mV to +734.4 mV -860 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 216
0.5 V 519 mV +37.24 mV to +38.76 mV 481 mV 3.4 V 3.4 V 3.419 V +37.24 mV to +38.76 mV 3.381 V 2.4 V 2.419 V +37.24 mV to +38.76 mV 2.381 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 217
-1.5 V -2.5 V -2.47 V +58.8 mV to +61.2 mV -2.53 V -1.5 V -1.47 V +58.8 mV to +61.2 mV -1.53 V -0.5 V -470 mV +58.8 mV to +61.2 mV -530 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 218
-560 mV 50 mV 150 mV +294 mV to +306 mV -150 mV +1 V 1.15 V +294 mV to +306 mV 850 mV -1 V -850 mV +294 mV to +306 mV -1.15 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 219
0.5 V 800 mV +588 mV to +612 mV 200 mV 3.4 V 3.4 V 3.70 V +588 mV to +612 mV 3.10 V 2.4 V 2.70 V +588 mV to +612 mV 2.10 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 220
-2.4 V -2.04 V +705.6 mV to +734.4 mV -2.76 V -1.5 V -1.5 V -1.14 V +705.6 mV to +734.4 mV -1.86 V -0.5 V -140 mV +705.6 mV to +734.4 mV -860 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 221
3 divisions above the center screen. NOTE. It’s important to use the Sense reading of the generator instead of the Output level when adjusting the generator to the specified voltage setting. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 222
Table 8: Gain accuracy on page 189.) ■ CHECK that the Difference of Measurement Mean is within the limits listed for the current vertical scale/termination/position/offset/ generator settings. Enter measurement mean difference value on test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 223
Test all channels: Repeat substeps 2.a through 2.i for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the generator output from the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 224
Select an unchecked channel: ■ From the tool bar, touch Measure and then Clear All to remove the previous measurement. ■ Press the Vertical button of the channel just confirmed to remove the channel from the display. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 225
-576.0 mV 50 mV +190 mV +372.40 mV to +387.60 mV -190 mV +0.25 V +690 mV +372.40 mV to +387.60 mV +310 mV -0.25 V -310 mV +372.40 mV to +387.60 mV -690 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 226
-576.0 mV 50 mV +190 mV +372.40 mV to +387.60 mV -190 mV +0.25 V +690 mV +372.40 mV to +387.60 mV +310 mV -0.25 V -310 mV +372.40 mV to +387.60 mV -690 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 227
-576.0 mV 50 mV +190 mV +372.40 mV to +387.60 mV -190 mV +0.25 V +690 mV +372.40 mV to +387.60 mV +310 mV -0.25 V -310 mV +372.40 mV to +387.60 mV -690 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 228
-576.0 mV 50 mV +190 mV +372.40 mV to +387.60 mV -190 mV +0.25 V +690 mV +372.40 mV to +387.60 mV +310 mV -0.25 V -310 mV +372.40 mV to +387.60 mV -690 mV MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 229
The DC test level should appear on screen. (If it does not return, the accuracy check has failed for the current vertical scale, position, and offset settings of the current channel.) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 230
Enter measurement mean difference value on test record. h. Repeat substeps 2.d through 2.g, using the next position, offset and generator settings listed in the table for the current vertical scale. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 231
Test all channels: Repeat substeps 2.a through 2.i for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the generator output from the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 232
Modify the default settings: ■ Set the Horizontal Scale to 1 ms/div. ■ Set the Record Length to 5000. ■ Set the Acquisition Mode to Hi Res. ■ Set the Trigger Source to Line. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 233
–12.5 mV to +12.5 mV +3.4 V +3.5 V +3.4 V +3.3737 V to +3.4263 V -3.4 V -3.5 V -3.4 V –3.4263 V to –3.3737 V Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 234
NOTE. It’s important to use the Sense reading of the generator instead of the Output level when adjusting the generator to the specified voltage setting. Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 235
Test all channels: Repeat substeps 2.a through 2.f for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the generator from the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 236
Select an unchecked channel: ■ From the tool bar, touch Measure and then Clear All to remove the previous measurement. ■ Press the Vertical button of the channel just confirmed to remove the channel from the display. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 237
■ Set the generator to the level and polarity indicated in the table for the vertical scale, position, and offset settings you have made. Set as precisely as the instrument's offset resolution permits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 238
Test all channels: Repeat substeps 2.a through 2.f for all four channels. 3. Disconnect the hookup: a. Set the generator output to 0 V. b. Disconnect the generator from the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 239
Hook up the test-signal source: Connect the sine wave output of a sine wave generator to Ch 1 through a power splitter. Connect the power sensor of the power meter to the power splitter. (See Figure 20: Analog MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 241
If your generator cannot output the required amplitude, determine its maximum output at the Test frequency, and use this for the reference amplitude. The -3 dB limit can be calculated as: 0.707 × reference amplitude. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 242
When doing substep 2.f, skip the subparts that turn on the Chx Amplitude mean measurement until you check a new channel. ■ When selecting a new channel and before doing substep 2.f, touch the Clear All button to remove the previous channel measurements. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 243
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series) Test all channels: Repeat substeps 2.a through 2.g for all four channels. 3. Disconnect the hookup: Disconnect the test hook up from the input connector of the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 244
Hook up the test-signal source: Connect the sine wave output of the sine wave generator to Ch 1 through a power splitter. Connect the power sensor of the power meter to the power splitter. Set the output of the MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 246
If your generator cannot output the required amplitude, determine its maximum output at the Test frequency, and use this for the reference amplitude. The -3 dB limit can be calculated as: 0.707 × reference amplitude. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 247
2.d. ■ All instruments except the MSO/DPO71254C and MSO/ DPO72504DX: Set the frequency of the generator, as shown on MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 248
Check remaining vertical scale settings against limits: ■ Check the remaining vertical scale settings for the channel under test by repeating substeps 2.a through 2.h for each of the remaining scale settings for the channel under test. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 249
Clear All button to remove the previous channel measurements. Test all channels: Repeat substeps 2.a through 2.i for all four channels. 3. Disconnect the hookup: Disconnect the test hook up from the input connector of the channel last tested. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 250
Measure the impedance: Read and record the measured impedance. d. Add the two measurements and divide the result by 2. e. Subtract the resistance of the multimeter leads from the average that you calculated. Enter the result on the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 251
Set the Vertical Scale of the channel to 10 mV (6.25 mv for >20 GHz models) per division. c. Repeat steps 2 through 3. 5. Disconnect the hookup: Disconnect the equipment from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 252
5. Set the Source Measure Unit to output –65 mA, with the overvoltage limit set to –3.5 V. Verify that the measured voltage is ≥–3.42 V. Enter the value in the test record. 6. Move the test setup to an unchecked input channel. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
10. Check the completion status: If any of the 4 checks on any of the 4 channels is outside the limits, return the instrument to Tektronix Service for repair. Time base system checks These procedures check those characteristics that relate to the time base system and are listed as checked under Warranted Characteristics in Specifications.
Page 254
■ CHECK that the count on the timer-counter is within limits. (See Long term sample rate, delay time, and internal reference accuracy in the test record.) ■ Enter the count on the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 255
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series) 3. Disconnect the hookup: Disconnect the equipment from the instrument. Figure 27: Timebase and delay time second test hookup MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 256
Touch Run SPC (Calibrate on some instruments) and wait for the signal path compensation to finish. b. Check the completion status: Enter the pass/fail status in the test record. If the Status is Fail, refer the instrument to qualified service personnel. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 257
Touch the Internal button to select the internal reference. c. Perform a signal path compensation: Touch Utilities and select Instrument Calibration. ■ ■ Touch Run SPC (Calibrate on some instruments) and wait for the signal path compensation to finish. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 258
90° right-angle female to male SMA adapter. The adapter is connected to one side of the female SMA T connector. The other side of the SMA T is connected to a 60 inch 50 Ω coaxial MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 259
From the button bar, touch Measure and select the Time tab to show the Time Measurements menu. ■ Touch the Pos Width button. ■ Touch Setups Statistics and then touch All. Touch Reset to reset the statistics. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 260
Press Run/Stop button to start the display. ■ Repeat step 1.d. ■ Touch Utilities, select User Preferences, select the Units tab, and then select the 1-2-3 button. 2. Disconnect all test equipment from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 261
From the button bar, touch the Horiz/Acq button. Select the Manual button to turn on Manual Mode. ■ Select the Acquisition tab. Press the IT button to turn on Interpolate Time Only. ■ From the button bar, touch the Vertical button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 263
Click one of the cursors to assign the multipurpose knobs to the cursors. ■ Using the multipurpose knobs, set the Cursor 1 x position to –5.0 ns, and set the Cursor 2 x position to +5 ns. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 264
Subtract the Max and Min statistic measurements. ■ ■ Enter the time on the test record. ■ The absolute value of the difference of the Max and Min measurements must be less than or equal to the pk-pk Delta-time MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
< 4 GHz models: Set the horizontal Scale to 2.5 ns. ■ ≥ 4 GHz models: Set the horizontal Scale to 2 ns. ■ < 4 GHz models: From the button bar, click the Vertical button; then click the Termination 50 Ω button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 266
254.) ■ Press the front-panel Advanced button and select the A Event tab; then pull down on Trigger Type and select Width triggering. ■ Pull down Pulse Width and select Inside limits. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 267
Model Limit < 4 GHz models 3.5 ns to 6.5 ns, inclusive ≥ 4 GHz models 4.77 ns to 5.23 ns, inclusive ■ Enter the time on the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 268
1.9 μs to 2.1 μs, inclusive ≥ 4 GHz models 1.9 μs to 2.1 μs, inclusive ■ Enter the time on the test record. 4. Disconnect the hookup: Disconnect the equipment from the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
0.35 dB of the reference frequency (10 MHz) through the trigger frequency being tested. Refer to the Sine Wave Generator Leveling Procedure if your sine wave generator does not have automatic output amplitude leveling. (See Sine wave generator leveling procedure on page 281.) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 270
From the button bar, touch Measure. ■ Touch Setup Ref Levs; then touch the Min-Max button. ■ Touch the Setup button and select the Ampl tab; then touch the Amplitude button. ■ Touch the X (close) button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 271
≥ 4 GHz models 4 divisions < 4 GHz models 7 divisions ■ Now fine adjust the generator output until the Ch 1 Amplitude readout indicates the amplitude is as follows (Readout may fluctuate): MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 272
Enter pass or fail in the test record. ■ Leave the trigger system triggered on the positive slope of the waveform before continuing to the next step. Figure 34: Measurement of trigger sensitivity - 50 MHz results shown MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 273
Press the Trigger Slope button to switch between trigger slopes. Use the Trigger Level knob to stabilize the trigger if required. ■ Enter pass or fail in the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 274
≥ 4 GHz models 200 mV DPO7354C 250 mV DPO7254C 600 mV DPO7104C 600 mV DPO7054C 600 mV ■ Disconnect the cable at Ch 1 and reconnect it to Ch 1 through an attenuator: MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 275
Press the Trigger Slope button to select the negative slope. Adjust the Trigger Level knob so that there is a stable trigger. ■ CHECK that the trigger is stable. ■ Enter pass or fail in the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 276
■ Set the generator frequency to full bandwidth as follows: Model Generator frequency ≥ 4 GHz models 1 GHz < 4 GHz models 250 MHz ■ Set the generator amplitude on screen as follows: MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
5. Disconnect the hookup: Disconnect the equipment from Aux In and the channel last tested. Output signal checks The procedure that follows checks those characteristics of the output signals that are listed as checked under Warranted Characteristics in Specifications. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
2. Confirm Aux Out is within limits for logic levels: a. Display the test signal: ■ Press the Vertical Ch 1 button to display that channel. ■ < 4 GHz models: Touch Vertical, select Vertical Setup, and then touch Termination 50 Ω. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 279
>20 GHz models: CHECK that the Ch 1 High readout is ≥0.5 volt and that the Ch 1 Low readout ≤0.125 volts. 3. Disconnect the hookup: Disconnect the test setup from the inputs and outputs. Figure 36: Measurement of trigger out limits MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Use the Vertical Position knob to center the display on screen. ■ From the button bar, touch Horiz/Acq and select the Acquisition tab. ■ Touch Average and set the number of averages to 128. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 281
Hook up the DC standard source: ■ Set the output of a DC calibration generator to off or 0 volts. ■ Connect the output of a DC calibration generator to Ch 1. Refer to the following figure. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 282
Press the X (close) button to remove the menus from the display. (See Figure 40: Measurement of probe compensator amplitude.) Figure 40: Measurement of probe compensator amplitude e. Check against limits: ■ Subtract the value just obtained (base level) from that obtained previously (top level). MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
One precision 50 Ω coaxial cable (Item 4) (See Prerequisites on page 116.) Also, the One sine-wave generator (Item 9) instrument must have passed Check DC Voltage One adapter (Item 19) Measurement Accuracy. (See Signal acquisition system checks on page 176.) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 284
From the button bar, touch Cursors and then the Setup button. If using the menu bar, touch Cursors and then select Cursor Setup. Touch the Cursor button to toggle it on and display the cursors. b. Set the Tracking Mode to Tracking. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 285
Touch the Edit and then the Clear button. o. Enter data into Serial Pattern Data field for the next setting in the table that is not yet checked. (See Table 19: Serial pattern data on page 272.) p. Touch Enter. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 286
Verify that the instrument triggers one Unit Interval (UI) after the 0 in the input signal. The absolute value of the T2 cursor readout must be ≤325 ps. Enter pass or fail in the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 287
Verify that the instrument triggers one Unit Interval (UI) after the 1 in the input signal. The absolute value of the T2 cursor readout must be ≤325 ps. Enter pass or fail in the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
The oscilloscope must meet the prerequisites. One sine-wave generator (Item 9) Also, the instrument must have passed Check One adapter (Item 19) DC Voltage Measurement Accuracy. (See Signal acquisition system checks on page 176.) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 290
Set the sine-wave generator to output one of the input frequencies in the table that is not yet checked, starting with the first setting. (See Table 21: Clock recovery input frequencies and baud rates on page 279.) MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
2. Set the Generator: ■ Set the sine wave generator to a reference frequency of 10 MHz. ■ Adjust the sine wave generator amplitude to the required number of divisions as measured by the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 294
Input the correction factor and/or the new frequency into the level meter. ■ Adjust the sine wave generator amplitude until the level meter again reads the value noted in step 3. The signal amplitude is now correctly set for the new frequency. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Set the sine wave generator to a reference frequency of 10 MHz. ■ Adjust the sine wave generator amplitude to the required number of divisions as measured by the instrument. Figure 47: Equipment setup for maximum amplitude MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 296
3. The signal amplitude is now correctly set for the new frequency. ■ Disconnect the sine wave generator from the power sensor. ■ Connect the sine wave generator to the instrument. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
The performance verification procedures verify the performance of your instrument. They do not adjust your instrument. If your instrument fails any of the performance verification tests, you should return the instrument to Tektronix for adjustment or repair. Test record...
Page 305
-895.3 mV BW, 50 Ω 100 mV/div 5.0 V 4.965 V 5.035 V 100 mV/div -5.0 V -5.035 V -4.965 V Use this value for both the calibrator output and the oscilloscope offset setting. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 306
-895.3 mV BW, 50 Ω 100 mV/div 5.0 V 4.965 V 5.035 V 100 mV/div -5.0 V -5.035 V -4.965 V Use this value for both the calibrator output and the oscilloscope offset setting. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 307
-99.5 V -100.2 V -98.80 V 3 V/div 99.5 V 98.4 V 100.6 V 3 V/div -99.5 V -100.6 V -98.4 V Use this value for both the calibrator output and the oscilloscope offset setting. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 327
4.22 V -0.1 V 0.1 V 3.78 V 4.22 V -0.1 V 0.1 V 3.78 V 4.22 V -0.1 V 0.1 V 3.78 V 4.22 V -0.1 V 0.1 V 3.78 V 4.22 V MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
97.) (The warm-up requirement is usually met in the course of meeting the Self Test prerequisites listed above). ■ Support sensor, probe, and adapter setups to avoid stress or torque when connected to the device under test (DUT). MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Wait. Signal-path compensation may take 5 to 15 minutes to run. d. Verify that the SPC Status is Pass and that the Temperature Status is Ready. 3. Return to regular service: Click the X (close) button to exit the instrument calibration control window. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
7. Repeat the tests at 250 kΩ as follows: a. Set the calibrator impedance to 1 MΩ. b. In the toolbar, click the Utilities menu and select Instrument Diagnostics. c. Click the on-screen Manual Product Verification button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
1. Attach a 50 Ω terminator to the oscilloscope channel 1 input. 2. Push the front-panel Default Setup button. 3. Set the oscilloscope impedance to 50 Ω as follows: a. Push the front-panel oscilloscope Vertical V menu button. b. Set the Termination to 50 Ω. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 332
Push the front-panel button to select the next channel to be tested. c. Move the 50 Ω terminator to the channel input to be tested. d. Starting from step 3, repeat the procedure until all channels have been tested. 15. Press the Menu Off button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Push the Trigger Menu front-panel button. b. In the Trigger menu, under Source, click Select. c. Click Line to set the trigger source to AC line, and then click OK. 7. Set the Vertical Scale to 2 mV/division. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 334
9. Record Gain Accuracy as follows: a. Calculate Vdiffas follows: Vdiff= | Vnegative-measured- Vpositive-measured | b. Enter Vdiffin the worksheet. (See Table 22: Gain expected worksheet page 322.) c. Calculate Gain Accuracy as follows: MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Move the DC voltage source connection to the channel input to be tested. d. Starting from step 3, repeat the procedure until all channels have been tested. 13. Press the Menu Off button. Check offset accuracy This test checks the offset accuracy. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 336
Push the front-panel Acquire button. b. In the Acquisition window, select the Acquisition tab at the left. c. Click Average Acquisition mode. d. Ensure that the number of averages is set to 16. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 337
Move the DC voltage source connection to the channel input to be tested. d. Starting from step 3, repeat the procedure until all channels have been tested. 17. Press the Menu Off button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
NOTE. At some V/div settings, the generator may not provide 8 vertical divisions of signal. Use the maximum output setting of the generator to obtain as many vertical divisions of signal as possible. 7. Set the Horizontal Scale to 1 ms/division. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 339
11. Record the peak-to-peak measurement as follows: a. With the Pk-Pk measurement still displayed from step 8, view the measurement at the new frequency. b. This will provide a mean V of the signal. Call this value V bw-pp MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 340
1 MΩ impedance as follows: a. Change the calibrator impedance to 1 MΩ. b. Push the front-panel Vertical V menu button. c. Set the Termination to 1 MΩ. d. Repeat steps 4 through 13. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Push the front-panel Wave Inspector Measure button. b. In the Measurement Setup window, select Statistics. c. Click Reset. 8. Read and make a note of the RMS Mean value. This is the Averaged Mean Value (AMV). MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Starting at step 4, repeat the procedure for each bandwidth setting in the test record. 13. Press the Menu Off button. Check sample rate and delay time accuracy This test checks the sample rate and delay time accuracy (time base). MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 343
80 ms. 12. Set the horizontal scale to 500 ns/div using the Horizontal Scale knob or by clicking in the Scale box and using the up/down arrows. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Click the Sample Rate field and set the sample rate to one of the unchecked values listed in the test record. e. Click the Record Length field to bring up the on screen keypad. Reduce the record length until the correct Horizontal Scale is set. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 345
Push the front-panel channel button for the next channel to be tested. c. Connect the signal source to the input for that channel. d. Repeat the procedure from step 5 until all channels have been tested. 11. Press the Menu Off button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Click in the Global Threshold box and set the value to 0.00 V. Click Apply. g. All thresholds are now set for the 0 V threshold check. 5. Set the Horizontal Scale to 5 μs/div. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 347
If the signal level is a logic high or is alternating between high and low, repeat this step (decrement Vs by 10 mV, wait 3 seconds, and check for a static logic low) until a value for Vs+ is found. 16. Push the front-panel Trigger Menu button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 348
Connect the DC voltage source to the digital channel to be tested. e. Starting from step 7, repeat the procedure until all 16 digital channels have been tested. 22. Press the Menu Off button. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Set the Termination to 50 Ω. c. Push the front-panel Wave Inspector Measure button. d. Ensure that the Amplitude tab is selected. e. Select the Low measurement. Enter the Low measurement reading in the test record. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Page 350
Performance verification (MSO/DPO5000/B series) g. Select the High measurement. h. Enter the High measurement reading in the test record. This completes the performance verification procedure. MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series...
Need help?
Do you have a question about the MSO70000C/DX and is the answer not in the manual?
Questions and answers