Replacing The Column-Outlet Tubing In A Cm-30S Installed In An Apc System - Waters ACQUITY CM-30S Overview And Maintenance Manual

30-cm single zone column manager
Table of Contents

Advertisement

3 Maintaining the 30-cm single-zone column manager
7.
Plug the leak sensor connector into the receptacle on the front of the
CM-30S.
8.
Power-on the CM-30S.
9.
In the instrument console, select Column Manager from the system tree.
10. In the column manager information window, click Control > Reset CM,
to reset the CM-30S.
Replacing the column-outlet tubing in a CM-30S installed
in an APC system
Warning:
and then allow the column compartment and its components to cool for
60 minutes before touching them. Monitor the column compartment's
internal temperature to ensure all components are cool.
Notice:
chemical-resistant, powder-free gloves when replacing the
passive-preheater assembly.
Required materials
Gloves: clean, powder-free, chemical-resistant
Replacement column-outlet tubing
76
October 7, 2014, 715004178 Rev. B
Leak sensor
connector receptacle
To prevent burn injuries, set the column temperature to Off,
To prevent contaminating system components, wear clean,