Kicker Replacement - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Kicker
Replacement
Thermo Fisher Scientific
Retaining Screws
(2 each heater)
Heaters
Figure 7–15. Replacing the Heater Assembly
Use the following procedure to replace the kicker assembly (Figure 7–16).
Equipment Required:
Kicker assembly
Cable ties
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the plumbing connections from the flow transducer, the
optical bench, and the rear bulkhead. Note the plumbing connections
to facilitate reconnection.
Model 43i Trace Level-Enhanced Instruction Manual
Servicing

Kicker Replacement

7-33