Lamp Intensity - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Min and Max Sample

Lamp Intensity

Min and Max Lamp
Intensity Limits
Thermo Fisher Scientific
The Minimum Sample Flow alarm limit screen is used to change the
Flow Limits
minimum sample flow alarm limit. The minimum and maximum sample
flow screens function the same way.
In the Main Menu, choose Alarms > Sample Flow > Min or Max.
The Lamp Intensity screen displays the current lamp intensity reading and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 20 to 100 percent. If the lamp intensity reading goes beyond
either the minimum or maximum limit, an alarm is activated. The word
"ALARM" appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Lamp Intensity.
The Minimum Lamp Intensity alarm limit screen is used to change the
minimum lamp intensity alarm limit. The minimum and maximum lamp
intensity screens function the same way.
In the Main Menu, choose Alarms > Lamp Intensity > Min or Max.
SAMPLE FLOW:
ACTUAL MIN:
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG
LAMP INTENSITY:
ACTUAL
>MIN
MAX
RANGE
AVG
LAMP INTENSITY:
ACTUAL MIN:
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG
Model 43i Trace Level-Enhanced Instruction Manual
0.250 LPM
0.300 LPM ?
DIAGS ALARM
90 %
40 %
100 %
DIAGS ALARM
40 %
50 % ?
DIAGS ALARM
Operation
Alarms Menu
3-61