to Table B–10. This command responds with "feature not enabled" if the
I/O expansion board is not detected.
Send:
analog iout range 4
Receive:
analog iout range 4 2
set analog iout range channel range
This command sets analog current output channel to the channel range
where channel is between 1 and 6 inclusive, and range is set according to
Table B–10. The example that follows sets current output channel 4 to the
0-20 mA range. This command responds with "feature not enabled" if the
I/O expansion board is not detected.
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
Table B–10. Analog Current Output Range Values
Range
1
2
0 [cannot be set to this, but may report]
analog vin channel
This command retrieves the analog voltage input channel data, both the
calculated value and the actual voltage. In the example that follows, the
"calculated" value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with "feature not enabled" if the I/O expansion board
is not detected.
Send:
analog vin 1
Receive:
analog vin 1 75.325 2.796
analog vout range channel
This command reports the analog voltage output channel range, where
channel is between 1 and 6 inclusive, according to Table B-11.
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range channel range
This command sets analog voltage output channel to the range, where
channel is between 1 and 6 inclusive, and range is set according to Table B–
11. The example that follows sets channel 2 to the 0-10 V range.
C-Link Protocol Commands
Output Range
0-20 mA
4-20 mA
Undefined
Model 43i Trace Level-Enhanced Instruction Manual
I/O Configuration
B-39