Reset User Calibration Default; Calibration Menu; Calibrate So Background - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation

Calibration Menu

Reset User Calibration
Calibration
CALIBRATION:
>CAL SO2 BACKGROUND
CAL SO2 COEFFICIENT
ZERO/SPAN CHECK
RANGE
Calibrate SO
Background
3-18
Model 43i Trace Level-Enhanced Instruction Manual
The Reset User Calibration Default screen allows the user to reset the
calibration configuration values to factory defaults.
Default
In the Main Menu, choose Calibration Factors > Reset User Cal Defaults.
RESTORE DEFAULT CAL:
RANGE
AVG
DIAGS ALARM
The Calibration menu is used to calibrate zero and span. The calibration
menu is similar for the single, dual, and autorange mode as shown below.
Menu
The dual and autorange modes have two SO
"LO"). This allows each range to be calibrated separately. This is necessary
if the two ranges used are not close to one another. For example, a low SO
range of 10 ppb and a high SO
about calibration, see Chapter 4, "Calibration".
In the Main Menu, choose Calibration.
AVG
DIAGS ALARM
The Calibrate SO
2
background, or perform a "zero calibration". Before performing a zero
calibration, ensure the analyzer samples zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, "Calibration".
In the Main Menu, choose Calibration > Cal SO
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT
PRESS
RANGE
range of 1000 ppb. For more information
2
CALIBRATION:
>CAL SO2 BACKGROUND
CAL HI SO2 COEFFICIENT
CAL LO SO2 COEFFICIENT
RANGE
Background screen is used to adjust the SO
2
RESTORE
TO CONFIRM RESTORE
AVG
DIAGS ALARM
span factors ("HI" and
2
AVG
DIAGS ALARM
2
Background.
2
Thermo Fisher Scientific
TO?x
2