Thermo Scientific 43i Instruction Manual page 269

Pulsed fluorescence so2 analyzer

Advertisement

Thermo Fisher Scientific
Selection
ppm
mg/m
7
100
250
8
C1
C1
9
C2
C2
10
C3
C3
custom range
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The example that follows reports that custom range 1 is defined to 5.50
ppm.
Send:
custom 1
Receive:
custom 1 550E-2 ppm
set custom range range value
set custom 1 range value
set custom 2 range value
set custom 3 range value
These commands are used to set the maximum concentration for any of the
three custom ranges 1, 2, or 3 to range value, where value is a floating-point
number representing concentration in ppm or mg/m
follows sets the custom 1 range to 55.5 ppm.
Send:
set custom 1 range 55500
Receive:
set custom 1 range 55500 ok
range mode
This command reports the current range mode.
Send:
range mode
Receive:
range mode single
set range mode mode
This command sets the current range mode to single, dual, or auto. The
example that follows sets the range mode to single.
Send:
set range mode single
Receive:
set range mode single ok
gas mode
This command reports the current mode of sample, zero, or span. The
example that follows reports that the gas mode is sample.
3
Model 43i Trace Level-Enhanced Instruction Manual
C-Link Protocol Commands
Measurement Configuration
3
. The example that
B-29