Figure B-1. Flags - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

C-Link Protocol Commands
Measurements
B-10
Model 43i Trace Level-Enhanced Instruction Manual
Send:
pres
Receive:
pres 753.4 mmHg
react temp
This command reports the current reaction chamber temperature. The
example that follows reports that the current reaction chamber temperature
is 45.2 °C.
Send:
react temp
Receive:
react temp 45.2 deg C
flags
This reports 8 hexadecimal digits (or flags) that represent the status of the
flash lamp, LED, pressure and temperature compensation status, gas units,
gas mode, and alarms. To decode the flags, each hexadecimal digit is
converted to binary as shown in the Table B–1. It is the binary digits that
define the status of each parameter. In the example that follows, the
instrument is reporting that the flash lamp is on, and the SO
concentration alarm is activated.
Send:
flags
Receive:
flags 40028000
Figure B–1. Flags
high
2
Thermo Fisher Scientific