Thermistor Replacement

Thermistor
Replacement
7-34
Model 43i Trace Level-Enhanced Instruction Manual
Kicker Ties
Kicker Bracket
Figure 7–16. Replacing the Kicker Assembly
3. Cut the ties securing the kicker assembly to the kicker bracket and
remove the kicker.
4. Install the new kicker, following the previous steps in reverse.
Use the following procedure to replace the ambient temperature thermistor
(Figure 7–17).
Equipment Required:
Thermistor
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Squeeze the thermistor latch and pull the thermistor from the AMB
TEMP connector on the measurement interface board.
3. Snap the new thermistor into the AMB TEMP connector.
Flow Transducer