Max Zero and Span Offset

Zero and Span Auto
Calibration
SO
Concentration
2
Thermo Fisher Scientific
ZERO CHECK:
ALARM:
RESPONSE:
>MAX OFFSET:
RANGE
The Max Zero Check Offset screen is used to change the maximum zero
check offset. The maximum zero and span offset screens function the same
way.
In the Main Menu, choose Alarms > Zero or Span Check > Max Offset.
MAX ZERO CHECK OFFSET:
CURRENTLY:
SET TO:
RANGE
The Zero Auto Calibration screen (read only) allows the user to view the
status of the most recent auto background calibration. The zero and span
auto calibration screens are visible only if the auto calibration option is
enabled and function the same way.
In the Main Menu, choose Alarms > Zero or Span Autocal.
ZERO AUTO CALIBRATION:
ALARM:
RESPONSE:
RANGE
The SO
Concentration screen displays the current SO
2
sets the maximum alarm limit. Acceptable alarm limits range from 0 to
1000 ppb or 2000 μ g/m
floor trigger (alarm is triggered when the concentration falls below the
minimum value) or a ceiling trigger (alarm is triggered when the
concentration goes above the minimum value). If the SO
AVG
DIAGS ALARM
5.0
000010.00 ?
MOVE CURSOR
SAVE
AVG
DIAGS ALARM
AVG
DIAGS ALARM
3
. The minimum alarm may be programmed as a
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Alarms Menu
OK
1.0
5.0
OK
1.2
concentration and
2
concentration
2
3-63