Download Print this page

Table B-6. Standard Ranges; Table B-7. Extended Ranges - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

C-Link Protocol Commands
Measurement Configuration
B-28
Model 43i Trace Level-Enhanced Instruction Manual
Send:
range
Receive:
range 6: 5000E-2 ppm
set range Selection
set high range Selection
set low range Selection
These commands select the SO
and Table B–7. The example that follows sets the SO
100 ppm.
Send:
set range 3
Receive:
set range 3 ok
Table B–6. Standard Ranges
Selection
ppm
mg/m
0
0.05
0.2
1
0.1
0.5
2
0.2
1
3
0.5
2
4
1
5
5
2
10
6
5
20
7
10
25
8
C1
C1
9
C2
C2
10
C3
C3
Table B–7. Extended Ranges
Selection
ppm
mg/m
0
0.5
2
1
1
5
2
2
10
3
5
20
4
10
50
5
20
100
6
50
200
full-scale ranges, according to Table B–6
2
3
3
full-scale range to
2
Thermo Fisher Scientific

Advertisement