Download Print this page

Hydrocarbon Kicker; Optics; Figure 8-1. Hardware Components - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

System Description
Hardware
Reaction Chamber Assy
Flow Transducer
Permeation Oven
Assy (Opt)
Pressure Transducer
Thermistor
Measurement Interface
Board
LCD Display
Front Panel Board

Hydrocarbon Kicker

8-2
Model 43i Trace Level-Enhanced Instruction Manual
Flash Intensity Assy
PMT Assy
Power Supply
Figure 8–1. Hardware Components
The hydrocarbon kicker removes hydrocarbons from the gas stream while
leaving the SO
permeation principle using differential pressure to force hydrocarbon
molecules to pass through the tube wall. The differential pressure is created
across the tube wall as sample gas passes through a capillary tube which
reduces its pressure. The sample gas is then fed to the shell side of the
hydrocarbon kicker. The kicker may be optionally mounted to the bench
to provide heating.
The optics section provides the light source for the fluorescence reaction

Optics

and optimizes the reaction with a system of lenses and mirrors. It includes a
flash lamp, condensing lens, mirror assembly, relay lens, and light baffle.
Mirror Housing Assy
Chamber Inlet Assy
HVPS
Input Board Assy
Optional Transformer
(not shown)
concentration unaffected. It operates on a selective
2
Flash Lamp Housing
Flash Trigger Assy
Kicker Assy
Solenoid Valves (Opt)
I/O Expansion
Board (Opt)
Digital Output Board
Mother Board
Thermo Fisher Scientific
Fan
Pump

Advertisement