Thermo Scientific 43i Instruction Manual page 257

Pulsed fluorescence so2 analyzer

Advertisement

Thermo Fisher Scientific
Send:
list stream
Receive:
list stream
field index variable
x x time
1 1 so2
2 18 pmtv
er xy
lr xy
sr xy
x = | 0 | 1 |
y = | 0 | 1 | 2 |
format" command)
These commands report the last long or short records stored and dynamic
data. In the example that follows, the command requests a long record with
no checksum, in ASCII format with text. For details on how to decode the
flag fields within these records, see the "flags" command.
Send:
lr01
Receive:
lr01
18:09 05-12-03 flags 5c000044 SO2 9658E-5 low SO2 8920E-5
Intt 28.6 Rctt 25.6 Pgast 0.00 Pres 720.3 Smpfl 0.489 PMt V
-808 LmpV 449 Lmpi 4740E+1
erec
This command returns a brief description of the main operating conditions
at the time the command is issued (i.e. dynamic data). The example that
follows shows a typical response.
The format is defined by the current settings of "format" and "erec format"
commands. For details on how to decode the flag fields within these
records, see the "flags" command.
Send:
erec
Receive:
erec
09:45 04-07-05 flags 5C040000 so2 0.001 4 loso2 0.001 4
pmtv 595.331 FlshV 797.227 Intt 29.263 Rctt 44.938 PGast
0.000 Smpfl 0.440 Pres 715.478 avgt 10 lo avgt 10 SO2bkg
6.795 SO2 coef 1.000 lo SO2 coef 1.000 SO2range 10000.000 lo
SO2range 10000.000 Lmpi 92.000
lrec
srec
lrec xxxx yy
srec xxxx yy
: Reply termination format (see "set format format"
command)
: Output format (see "set erec/lrec/srec format
Model 43i Trace Level-Enhanced Instruction Manual
C-Link Protocol Commands
Datalogging
B-17