Analog Output Calibrate
Full-Scale
Analog Input
Calibration
Thermo Fisher Scientific
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a meter to the output and adjust output until it reads the value shown in
the set output to: field.
In the Main Menu, choose Service > Analog Output Calibration > Selected
Channel > Calibrate Full Scale.
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
The Analog Input Calibration menu is a selection of analog input to
calibrate, and allows the user to select the calibration action zero or span.
The analog input calibration is visible only when the instrument is in
service mode. For more information on the service mode, see "Service
Mode" earlier in the chapter.
Note This screen is present only when the I/O expansion board is
installed.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Analog Input Calibration > Input
Channel 1-8.
SPAN
3697
INC/DEC
10 V
AVG
DIAGS ALARM
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
V1
3-73