Operation
Instrument Controls Menu
Digital Input Settings
3-40
Model 43i Trace Level-Enhanced Instruction Manual
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS
ZERO MODE
SPAN MODE
RANGE
SAMPLE MODE
SO2 MODE
PURGE MODE
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the instrument parameter or logic
state to change for the relay selected.
Note The digital inputs must be asserted for at least one second for the
action to be activated.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings.
DIGITAL INPUT SETTINGS:
>1
NOP
BACKGROUND
2
NOP
3
NOP
4
NOP
5
NOP
SPAN
6
NOP
7
NOP
RANGE
8
NOP
9
NOP
10
NOP
11
NOP
12
NOP
13
NOP
14
NOP
15
NOP
16
NOP
AVG
DIAGS ALARM
SET
CAL TO LO SPAN
AOUTS TO ZERO
AOUTS TO FS
CAL TO HI
NONE
NONE
AVG
DIAGS ALARM
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
NONE
<--