C-Link Protocol Commands
Measurements
B-8
Model 43i Trace Level-Enhanced Instruction Manual
Send:
avg time
Receive:
avg time 11:300 sec
set avg time selection
set high avg time selection
set low avg time selection
These commands set the averaging time, high and low averaging times,
according to Table B–2. The example that follows sets the low range
averaging time to 120 seconds.
Send:
set low avg time 8
Receive:
set low avg time 8 ok
Table B–2. Averaging Times
Selection
Averaging Time (seconds)
0
1 seconds
1
2
2
5
3
10
4
20
5
30
6
60
7
90
8
120
9
180
10
240
11
300
so2
high so2
low so2
These commands report the measured SO
in single range, or high and low SO
mode. The example that follows shows that the SO
ppm.
Send:
so2
Receive:
so2 0040E+0 ppm
concentration when operating
2
when operating in dual or auto range
2
concentration is 40
2