Table B-12. Default Output Assignment - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Thermo Fisher Scientific
Send:
dout 4
Receive:
dout 4 11 GEN ALARM open
set dout channel index state
This command assigns digital output channel to be assigned to the action
associated with index, and assigns it an active state of state (open or closed).
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa channel
This reports the outputs of the 6 or 12 Digital to Analog converters,
according to Table B–12. The example that follows shows that the D/A #1
is 97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
Table B–12. Default Output Assignment
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
7
Current Output
8
Current Output
9
Current Output
10
Current Output
11
Current Output
12
Current Output
Single Range
Dual Range
Low SO
Low SO
2
2
Low SO
High SO
2
2
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Low SO
Low SO
2
2
Low SO
High SO
2
2
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Not Assigned
Model 43i Trace Level-Enhanced Instruction Manual
C-Link Protocol Commands
I/O Configuration
Autorange
High/Low SO
2
Range Status
Not Assigned
Not Assigned
Not Assigned
Not Assigned
High/Low SO
2
Range Status
Not Assigned
Not Assigned
Not Assigned
Not Assigned
B-41