Thermo Scientific 43i Instruction Manual page 273

Pulsed fluorescence so2 analyzer

Advertisement

Thermo Fisher Scientific
set date mm-dd-yy
mm = month
dd = day
yy = year
This command sets the date of the analyzer's internal clock. The example
that follows sets the date to December 1, 2004.
Send:
set date 12-01-04
Receive:
set date 12-01-04 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send:
set default params
Receive:
set default params ok
lamp status
This command reports the status of the flash lamp on or off. The example
that follows reports the flash lamp is on.
Send:
lamp status
Receive:
lamp status on
set lamp onoff
These commands set the flash lamp on or off. The example that follows
turns the flash lamp off.
Send:
set lamp off
Receive:
set lamp off ok
pmt status
This command reports the status of the PMT on or off. The example that
follows reports that the PMT is on.
Send:
pmt status
Receive:
pmt status on
set pmt onoff
These commands set the PMT on or off. The example that follows turns
the PMT off.
C-Link Protocol Commands
Model 43i Trace Level-Enhanced Instruction Manual
Hardware Configuration
B-33