Operation

Diagnostics Menu

Diagnostics

Program Versions

3-50
Model 43i Trace Level-Enhanced Instruction Manual
The Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The
Menu
analog input readings and analog input voltages are only displayed if the
I/O expansion board option is installed.
In the Main Menu, choose Instrument Controls > Diagnostics.
The Program Versions screen (read only) shows the version numbers of the
programs installed. Prior to contacting the factory with any questions
regarding the instrument, please note the product model program version
numbers.
In the Main Menu, choose Diagnostics > Program Versions.

Voltages

The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board item is only
displayed if the I/O expansion board is installed.
DIAGNOSTICS:
>PROGRAM VERSIONS
VOLTAGES
TEMPERATURES
PRESSURE
FLOW
LAMP INTENSITY
OPTICAL SPAN TEST
RANGE
AVG
ANALOG INPUT READINGS
ANALOG INPUT VOLTAGES
DIGITAL INPUTS
RELAY STATES
TEST ANALOG OUTPUTS
INSTRUMENT CONFIGURATION
CONTACT INFORMATION
PROGRAM VERSIONS:
PRODUCT:
MODEL 43iTLE
VERSIONS: 01.02.22.104
FIRMWARE: 07.92.19
RANGE
AVG
DIAGS ALARM
DIAGS ALARM