Analog Output Calibrate Zero - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation
Service Menu
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
RANGE
CURRENT CHANNEL 2
CURRENT CHANNEL 3
CURRENT CHANNEL 4
CURRENT CHANNEL 5
CURRENT CHANNEL 6
Analog Output Calibrate
3-72
Model 43i Trace Level-Enhanced Instruction Manual
service mode. For more information on the service mode, see "Service
Mode" earlier in the chapter.
Notes This adjustment should only be performed by an instrument service
technician.
Current channels are visible only if the I/O expansion board is installed.
In the Main Menu, choose Service > Analog Output Calibration > Voltage
Channel 1-6 or Current Channel 1-6.
AVG
DIAGS ALARM
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a meter
Zero
to the output and adjust the output until it reads 0.0 V on the meter.
In the Main Menu, choose Service > Analog Output Calibration > Selected
Channel > Calibrate Zero.
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
ANALOG OUTPUT CAL:
CONNECT METER TO OUTPUT!
SELECTED OUTPUT:
SET TO:
SAVE VALUE
SET OUTPUT TO:
RANGE
AVG
AVG
DIAGS ALARM
ZERO
V1
200
INC/DEC
0.0 V
DIAGS ALARM