Calibration; Table B-5. Stream Time Values - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Calibration

Thermo Fisher Scientific
set stream per number value
number value = | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to number value in seconds. The example that follows sets the
number value to 10 seconds.
Send:
set stream per 10
Receive:
set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to Table B–5.
Send:
stream time
Receive:
stream time 0
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to Table B–5. The example that follows
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5. Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal so2 coef
set cal high so2 coef
set cal low so2 coef
These commands will auto-calibrate SO
gas concentrations. The high and low commands are only available in dual
and auto range mode. If the mode is incorrect, the instrument responds
with "can't, wrong settings". The example that follows shows a successful
auto-calibration of the low SO
Send:
set cal low so2 coef
Receive:
set cal low so2 coef ok
coefficients based on SO
2
coefficient.
2
Model 43i Trace Level-Enhanced Instruction Manual
C-Link Protocol Commands
Calibration
span
2
B-23