Resistor; Calibrate Oven Thermistor; Permeation Oven Setpoint - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation
Service Menu
Calibrate Oven
Permeation Oven
3-76
Model 43i Trace Level-Enhanced Instruction Manual

Resistor

The Calibrate Gas Thermistor Resistance screen allows the user to view and
set the permeation oven gas thermistor resistance to a known resistor value.
The Calibrate Oven Thermistor menu is used to view and set the
permeation oven thermistor resistance to a known resistor value.
Thermistor
In the Main Menu, choose Service > Perm Oven Settings > Cal Oven
Thermistor.
The Permeation Oven Setpoint screen is used to select the permeation oven
state to not present, or the desired temperature choice of 30, 35, and 45
Setpoint
°C.
In the Main Menu, choose Service > Perm Oven Settings > Perm Oven
Selection.
CAL GAS THERM (BATH):
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
CAL GAS THERM (RESISTOR):
CURRENTLY:
SET TO :
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
CAL OVEN THERM
(RESISTOR):
CURRENTLY:
SET TO :
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
o
45.80
C
o
45.00
C
SAVE
DIAGS ALARM
3850 Ohms
04000
Ohms
SAVE
DIAGS ALARM
3850 Ohms
04000
Ohms
SAVE
DIAGS ALARM
Thermo Fisher Scientific