Permeation Oven Settings; Calibrate Gas Thermistor; Water Bath - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Permeation Oven

Calibrate Gas Thermistor

Thermo Fisher Scientific
The Permeation Oven Settings menu is used for setting up and calibrating
the permeation oven option. The permeation oven settings menu is visible
Settings
only when the instrument is in service mode. For more information on the
service mode, see "Service Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Perm Oven Settings.
The Calibrate Gas Thermistor submenu is used for calibrating the
permeation oven gas thermistor using a water bath or known resistance.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor > Water Bath or Known Resistor.

Water Bath

The Calibrate Gas Thermistor Water Bath screen allows the user to view
and set the permeation oven gas temperature to a known temperature
value.
PERM OVEN SETTINGS:
>CAL GAS THERMISTOR
CAL OVEN THERMISTOR
PERM OVEN SELECTION
FACTORY CAL GAS THERM
FACTORY CAL OVEN THERM
RANGE
AVG
CALIBRATE GAS THERMISTOR:
>WATER BATH
KNOWN RESISTOR
RANGE
AVE
Model 43i Trace Level-Enhanced Instruction Manual
DIAGS ALARM
DIAGS ALARM
Operation
Service Menu
3-75