Contact Information; Alarms Menu - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Contact Information

Alarms Menu

Thermo Fisher Scientific
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD
ZERO/SPAN VALVES
PERM OVEN
DILUTION
AUTO CALIBRATION
RANGE
The Contact Information screen displays the customer service information.
In the Main Menu, choose Diagnostics > Contact Information.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB:
RANGE
The alarms menu displays a list of items that are monitored by the analyzer.
If the item being monitored goes outside the lower or upper limit, the
status of that item will go from "OK" to either "LOW" or "HIGH",
respectively. If the alarm is not a level alarm, the status will go from "OK"
to "FAIL". The number of alarms detected is displayed to indicate how
many alarms have occurred. If no alarms are detected, the number zero is
displayed.
To see the actual reading of an item and its minimum and maximum
limits, move the cursor to the item and press
Items displayed are determined by the options displayed. The zero/span
check and auto calibration screens are visible only if the zero/span check or
auto calibration options are enabled. The motherboard status, interface
board status, and I/O expansion board status (if installed) indicates that the
power supplies are working and connections are successful. There are no
setting screens for these alarms.
In the Main Menu, choose Alarms.
AVG
DIAGS ALARM
WWW.THERMO.COM
AVG
DIAGS ALARM
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Alarms Menu
YES
YES
NO
YES
YES
.
3-57