Download Print this page

Hardware Configuration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Hardware
Configuration
Thermo Fisher Scientific
pres comp
This command reports whether pressure compensation is on or off. The
example that follows shows that pressure compensation is on.
Send:
pres comp
Receive:
pres comp on
set pres comp onoff
These commands turn the pressure compensation on or off. The example
that follows turns pressure compensation off.
Send:
set pres comp off
Receive:
set pres comp off ok
temp comp
This command reports whether temperature compensation is on or off.
The example that follows shows the temperature compensation is off.
Send:
temp comp
Receive:
temp comp off
set temp comp onoff
These commands turn the temperature compensation on or off. The
example that follows turns temperature compensation off.
Send:
set temp comp off
Receive:
set temp comp off ok
contrast
This command reports the screen's level of contrast. The example that
follows shows the screen contrast is 50%, according to Table B–8.
Send:
contrast
Receive:
contrast 5:50%
set contrast level
This command sets the screen's level of contrast, according to Table B–8.
The example that follows sets the contrast level to 50%.
Send:
set contrast 5
Receive:
set contrast 5 ok
C-Link Protocol Commands
Model 43i Trace Level-Enhanced Instruction Manual

Hardware Configuration

B-31

Advertisement